INSTRUCTIONS FOR USE
DA 170, DA 130
UM_EN_DA170_DA130_2018-09_ver2.6
28
/
37
After several seconds the LED bargraph indicator starts blinking between 50% and
100%.
This means that you are in the mode of reaction of disinfecting solution
Note
Due to the fact that syringe is controlled by autonomous valve, you have to perform filling of its hose manually:
Put the tip of the syringe to the tools holder, press the left button and hold until the disinfection solution flows out
from the tip. Leave the syringe in the hoses holder. Same procedure applies when rinsing with water after the
disinfection.
Switch the depressurasing switch to position „0“ or turn the dental unit off with the main switch. Let the disinfection
solution react whole night or at least 60 minutes.
Phase 2
– Rinsing of hoses
If you decided to stop the disinfection process, then:
•
Take off the disinfection solution bottle and replace it with distilled water bottle (system is depressurized from stage
1)
•
Pressurize the system: Switch the three-position toggle switch to position
"DESTIL" or turn on the
dental unit with the main switch. The bottle will pressurize (If the dental unit
was turned off then during
turning on the program will automatically return to the mode of reacting of disinfecting solution). Make sure the hoses
are correctly placed in the hose holder in the rinsing bowl.
•
Press button
and hold it at least for 4 seconds. System will shortly beep 3 times, the green
LED starts shining and the rinsing starts. Rinsing progress is indicated by continual shinning up of
LED diodes on the LED bargraph.
•
End of stage 2 (rinsing of hoses) and end of the program of disinfection are indicated by:
•
Green LED diode lights off
•
System beeps 3 times
•
PROGRAM IS AUTOMATICALLY FINISHED.
•
System returns to the standard user mode.
•
Step by step take out the hoses, fit in the hand-pieces and put the tools back to working positions.
Due to the fact that syringe is controlled by autonomous valve, its rinsing is necessary to be performed manually.
Put the syringe over the tools holder, press the left button and hold until clear water goes out from the tip (hold it
for 2 minutes).
After the syringe rising you can position it back to its working position on the dentist’s table.
Troubleshooting
Switch Off or Power Off of the Unit.
If accidental Switch Off or Power Off happens during one of these stages:
•
During hoses Filling in with disinfection solution
•
During disinfection solution reacts in the hoses
•
During hoses rinsing with water
then after Switching-On or Powering-On, triple beeping sound occurs and the system automatically loads itself into the
program mode of reacting of disinfecting solution in the hoses (Stage 2).
If the interruption happened during the hoses filling with disinfecting solution (stage 1) then close the program
by pressing button
and start it again.