INSTRUCTIONS FOR USE
DA 170, DA 130
UM_EN_DA170_DA130_2018-09_ver2.6
27
/
37
Disinfection Procedure
Preparation of Disinfection
•
Prepare a bottle for disinfecting solution (A), fill it with the disinfecting solution (B)
•
Open the door of the spittoon block
•
If the unit is equipped with a depressurized bottle or a central water supply accessories switch
the three-position switch in the spittoon block to the position "0" - depending on the unit
equipment. Otherwise, turn off the main switch of the dental unit to ensure depressurization.
•
Remove the bottle with distilled water
•
Screw the bottle that contains disinfecting solution
•
Set three-position switch to the position "DESTIL" or
turn on the dental unit.
•
The bottle will pressurize.
•
Close the door of the spittoon block
•
Place the holder of the hoses for tools to a bowl. (C)
Program control
The program is controlled from the main keyboard using the buttons : plus(+) and minus (-).
Status indication or the process stage is signalized also
Through the main keyboard and through the buzzer..
Starting the Disinfection
Before starting the program, the tools must be in their positions on the dentist’s
table. Needle valves must be open to the maximum to regulate the pressure of
waterlines.
Start the program by pressing both buttons
and hold (the system will beep 6 times)
until all LEDs are shining green.
You are in disinfection mode.
Step by step take out all the hoses and put them to the hoses holder.
If any tool remains in its position on the dentist’s table, you are at risk of being
splashed by disinfecting solution.
Phase 1
– Filling of hoses with solution
By pressing button
you start hoses filling with disinfection solution.
System will beep 3 times and the red LED shines up
– stage 1. has started.
The ongoing process of hoses filling is indicated on the LED bargraph
– green LEDs gradually shine up
When bargraph indicator shines at 100%, the hoses are filled in. System is automatically switched to
the mode of reacing of disinfection.
Start the program (transition to a
disinfection program)
Simultaniously
press Plus a
Minus
Stage 1 - Starting the filling of hoses
with disinfecting solution and
disinfection reaction
Press Plus
Stage 2 - Starting rinsing of hoses
with water
Press again
Plus
Completion of the program (leaving
the disinfection program)
Press Mínus
Position of the
switch to pressurize
the bottle