ASE 150 Operator’s Manual
84
Doc. 065207-02 9/08
7. Turn on the ASE 150 main power switch.
8. Reset the
PUMP STROKE
counter on the
EXTRACTION COUNTERS
screen (see
5.7
Replacing the Pressure Relief Valve
1. Follow the instructions in
to disconnect the gas source and
remove the right-side panel.
2. The pressure relief valve is installed in the upper left corner of the component
). Disconnect the yellow gas tubing (
) from the
elbow press fitting on the valve.
To disconnect a press fitting, use your fingers (or a small open-end wrench) to
press the ring on the fitting
in
, while at the same time pulling the tubing
out
.
3. Use a 1/4-inch open-end wrench to disconnect the stainless steel line (
)
from the right side of the valve. (This tubing connects to the pump
transducer.)
4. Use the wrench to disconnect the Hastelloy® line (
) from the left side of
the valve. (This tubing connects to the upper AutoSeal tip.)
Figure 5-18. Connections to the Pressure Relief Valve
1
1
2
2
3
3
4
4
Pressure Relief Valve
Hastelloy Line
Green PEEK Waste Line
Stainless Steel Line
Yellow Gas Tubing
Allen Hex Screw
2
4
5
3
1
6
1
2
3
4
5
6
7
7
Splash Guard
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...