ASE 150 Operator’s Manual
60
Doc. 065207-02 9/08
Error 009
Pump volume limit exceeded.
Cause:
An empty (or incorrectly installed) solvent reservoir.
Action:
Refill the solvent reservoir, if necessary. Check that the solvent
reservoir is correctly installed (see
Cause:
A leak somewhere in the system.
Action:
Remove the right-side panel of the ASE 150 (see
Inspect the following fittings for leaks (see
): pump
fittings, pump check valves, pressure transducer fittings, pressure
relief valve fittings, static valve fittings, and solvent line fittings. Use
a 1/4-inch open-end wrench (P/N 049452) to tighten any leaking
fittings.
Do not overtighten!
If tightening does not stop a leak, the tubing and fitting assembly
must be replaced. Contact Dionex for assistance.
Action:
Pull out the drip tray holder below the oven and check the drip tray. If
the tray contains liquid, remove it from the holder, dispose of the
liquid, and reinstall the tray. Then, remove the collection vessel and
waste bottle from the system. Remove the lower trim panel and
check for leaks near the oven and the AutoSeal mechanism.
Error 010
Analog-to-digital converter failed.
Cause:
Failure of the main PC board (printed circuit board). The board must
be replaced.
Action:
Contact Dionex for assistance.
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...