Dinel
®
9 .
o
peraTion
anD
seTTing
Set the level meter using 3 buttons placed on the display module
(see Chapter Set-up elements). After 5 min. of inactivity, the level
meter automatically returns back to the measurement mode. If the
password is active, the level meter will be also locked. The val-
ues that have not been confirmed using the button
will not be
saved! After the meter is locked, you cannot change the setting!
When you attempt to edit, the words "NO PASSWORD" will appear
on the display. How to unlock the level meter is given on page 16.
After connection of the supply voltage to the level meter the display
shows the logo "Dinel" and the text "Starting" (approx. 15 s). Then,
the level meter goes to the measuring mode and the display shows
the current measured value.
MIN LEVEL and MAX LEVEL
You can freely define the
minimum / maximum distance
from the front surface of the level meter
(item "LEVEL"), and assign to this distance any current (item "OUTPUT") at the output in the range
of 4 to 20 mA. The "DISPLAY" is intended to set the value displayed on the display. Setting the units
is done in the "UNITS".
For clarity it is advised to assign a minimum
level to current 4 mA and the maximum level
to current 20 mA.
ACTUAL LEVEL: Actual distance to level
OUTPUT: Definition of minimum / maximum
current in mA
LEVEL: Definition of the min / max level
DISPLAY: The value showed on the display
M I N L E V E L
M A X L E V E L
U N I T S
D A M P I N G
U N I T S
L E V E L :
m m
D I S P L A Y :
%
T E M P E R A T U R E
°C
S A V E D
If in the bottom of the display appears (when entering the values) the inscription "OUT OF LIMITS",
the value specified for the item "LEVEL" is outside the measuring range of the level meter. If the
inscription "SPAN TOO SMALL" is shown, it must be specified a larger span between Min and Max
values. For more information, see chapter "Specifications".
The decimal point position of the item 'LEVEL' is firmly set (according to the selected units), in the
item "DISPLAY" it is freely adjustable
9 .1 . B
asic
configuration
After the first start of the level meter it is necessary to perform the
basic configuration (setting of the measuring range, choice of units
and possibly damping). The settings are accessible in the basic
menu by pressing
the "BASIC SETTINGS".