Advanced Force Gauge User’s Manual
29
If the current % offset is as shown below, follow the
appropriate action.
Lower Value
Action
-4.9% to 4.9%
Normal offset provided
upper and lower values are
within 5% of each other
-5.0% to -9.9%
Contact Quantrol distributor
5% to 9.97%
for recalibration
-10% and greater
Contact Quantrol distributor
10% and greater
for loadcell replacement
These values are given as an indicator only – the
need for calibration/repair may vary according to the
individual characteristics of the load cell.
Press and hold
ESC
until you return to page 2 of main
menu.
A constant of multiplication can be applied to the load
values in the main display. This is useful for applica-
tions where you wish to convert the load reading into
a non-standard unit of measurement e.g. for co-
efficient of friction tests.
Press
ENTER
to select
xCONSTANT
. Press
UP
or
DOWN
to adjust the constant of multiplication. Press
ENTER
to confirm.
The contrast of the display may be increased and
decreased in this menu.
From page 2 of main menu press
UP
or
DOWN
keys
to position the arrow key on
CONTRAST
. Press
ENTER
to select
CONTRAST
menu and 1.234 will be
displayed. Press
UP
or
DOWN
to adjust the contrast
of the display and
ENTER
to confirm the setting. You
will now return to page 2 of the main menu.
x/CONSTANT
CONTRAST
Содержание Quantrol AFG 10
Страница 1: ...Advanced Force Gauge AFG AFTI User s Manual ...
Страница 4: ...Advanced Force Gauge User s Manual 4 ...
Страница 34: ...Advanced Force Gauge User s Manual 34 Stand ...
Страница 35: ...Advanced Force Gauge User s Manual 35 Freeze Drop Average Time Rate ...
Страница 36: ...Advanced Force Gauge User s Manual 36 Footswitch1 Footswitch2 Comms Constant ...
Страница 37: ...Advanced Force Gauge User s Manual 37 Information Calibration Contrast ...
Страница 38: ...Advanced Force Gauge User s Manual 38 ...
Страница 39: ...Advanced Force Gauge User s Manual 39 ...