Advanced Force Gauge User’s Manual
26
This function selects the display throughput rate, i.e.
the amount of averaging performed by the internal
electronics before the load reading is displayed.
There are three levels HI, MED and LO.
HI
Display updates quickly with little data
averaging (2000 Hz)
MED
Gauge default. (10 Hz)
LO
Maximum data averaging before values are
displayed. Smoothes ‘noisy’ data. (2 Hz)
To set
RATE
, press and hold the
MENU
key until
page 1 of the main menu appears. Using
UP
and
DOWN
move the arrow key to
RATE
.
Using
UP
and
DOWN
select the relevant level (
HI
,
MED
or
LO
) and press the
ENTER
key.
Press
ESC
to return to page 1 of the main-menu.
Figure 12
Main menu, page 2
From page 1 of main-menu press MENU to scroll to
page 2 of the main menu. See Figure 12.
The AFG has two footswitch input pins on the 15-pin
D connector. This allows a footswitch to be assigned
to perform any of the five main key functions,
MAX
,
UNITS
,
TXD
,
ZERO
and
RESET
.
This feature is useful when integrating the AFG into
test or production systems.
To assign the function of a key to FOOTSWITCH 1
use
UP
and
DOWN
to move the arrow key to
FOOTSWITCH 1
. Using
UP
and
DOWN
select the
relevant key (
MAX
,
UNITS
,
TXD
,
ZERO
or
RESET
)
and press the
ENTER
key. Press
ESC
to return to
page 2 of the main-menu.
RATE
FOOTSWITCH 1
Содержание Quantrol AFG 10
Страница 1: ...Advanced Force Gauge AFG AFTI User s Manual ...
Страница 4: ...Advanced Force Gauge User s Manual 4 ...
Страница 34: ...Advanced Force Gauge User s Manual 34 Stand ...
Страница 35: ...Advanced Force Gauge User s Manual 35 Freeze Drop Average Time Rate ...
Страница 36: ...Advanced Force Gauge User s Manual 36 Footswitch1 Footswitch2 Comms Constant ...
Страница 37: ...Advanced Force Gauge User s Manual 37 Information Calibration Contrast ...
Страница 38: ...Advanced Force Gauge User s Manual 38 ...
Страница 39: ...Advanced Force Gauge User s Manual 39 ...