EDXtreme User Instructions
9 Troubleshooting
Problem
Possible Cause
Solution
EDX powers on momentarily
and turns off
Low battery
Bad keypad
Replace with high quality alkaline batteries. Do not use rechargeable
batteries.
Have unit serviced.
EDX does not power on
Low battery
Bad keypad
Replace with high quality alkaline batteries. Do not use rechargeable
batteries.
Have unit serviced.
Batteries installed
backwards or no spring
contact
Insure that positive terminals of both batteries (nub) face inward –
towards the black cap. Check that spring is attached to the battery cap.
Software reset
Remove battery cap & reinstall after one minute. Attempt to turn power on
again.
Display contrast too light Hold the Right Arrow key down while pressing the F2 key several times to
increase the display contrast. If nothing occurs, release both keys. Press
the power button and try again.
EDX PSU Power Supply Remove Power Supply and install batteries to see if the EDX or
Communicator II will power up. Return EDX PSU to factory for repair/
troubleshooting.
Display is completely dark
Display contrast too dark Hold the Arrow key down while pressing the F1 key several times to
decrease the display contrast.
EDXtreme does not appear
accurate
Check installation &
system
Insure that shackles are in good working condition and aligned straight.
Verify system is applying force directly through the dynamometer with no
off center or torsional loads being applied to the instrument.
Local gravitational
variances
If being compared against dead-weights, check your local gravitational
constant. Use custom units to compensate or calibrate on-site.
Check repeatability
Place EDX in low-resolution mode. Lift an arbitrary weight several times
as close to capacity as possible. Record each weight reading. Do the
readings differ from each other? Calculate the standard deviation of the
readings using a spreadsheet such as Microsoft Excel. See if the
deviation is greater than 0.1% of the instrument capacity.
Compare against a
reference load.
Place EDX in low-resolution mode. Apply a known load near instrument
capacity. Check calibration date.
Radio communication
intermittent
Low batteries. Distance
is excessive or dead-
radio pocket
Bring remote closer to dynamometer. Allow several seconds to retrain.
Excessive radio noise or
interference in
environment
Remove dynamometer and remote from the environment. Attempt
communications in an area free of local radio signals.
Содержание EDXtreme
Страница 1: ...EDXtreme Dynamometer and Crane Scale User Instructions AWT35 501242 Issue AC Optional Communicator ...
Страница 10: ...8 EDXtreme User Instructions 1 8 Declaration of Conformity EDX and Communicator II ...
Страница 11: ...EDXtreme User Instructions 9 1 9 Declaration of Conformity EDX PSU ...
Страница 12: ...10 EDXtreme User Instructions ...
Страница 34: ...32 EDXtreme User Instructions 4 1 6 Setpts Future Feature ...
Страница 63: ...EDXtreme User Instructions 61 10 6 Good Force Measurement Practice ...
Страница 67: ......