30
EDXtreme User Instructions
4.1.3 Setup--About
The next soft key is the About. The About menu shows an assortment of information
about your Dillon instrument. This can be handy for maintaining calibration,
troubleshooting or determining if the firmware can be upgraded. Press this and access
the following soft key set (refer to
Figure 4.1
):
Device
Press this to see software revision and dynamometer information.
Calib
This soft key access the following soft key set:
Points
- Press this key to display the calibration loading points. This is
useful information that can be keyed in manually in the event instrument
memory is ever accidentally overwritten or corrupted.
Re-Cal-
Will show next re-calibration date and contact detail.
Print-
Press this key to transmit all the calibration information available to
a serial printer or computer.
O.Load
Lists the number of overloads that have occurred since the last calibration.
On Time-
Actual hours when unit is on.
Zero
Lists the current zero point compared to the calibration zero point. If the
zero point has moved significantly, this may indicate a serious overload
has occurred and the instrument should be returned for service.
Ntwrk
Lists radio information; the enabled channel and identifier.
Dillon suggests that calibration data be recorded and saved.
Содержание EDXtreme
Страница 1: ...EDXtreme Dynamometer and Crane Scale User Instructions AWT35 501242 Issue AC Optional Communicator ...
Страница 10: ...8 EDXtreme User Instructions 1 8 Declaration of Conformity EDX and Communicator II ...
Страница 11: ...EDXtreme User Instructions 9 1 9 Declaration of Conformity EDX PSU ...
Страница 12: ...10 EDXtreme User Instructions ...
Страница 34: ...32 EDXtreme User Instructions 4 1 6 Setpts Future Feature ...
Страница 63: ...EDXtreme User Instructions 61 10 6 Good Force Measurement Practice ...
Страница 67: ......