Y-axis Motor
• Function
1. Y-axis motor works on it as the slow axis scan.
2. Separate SD mode with HD mode as the slow scan speed.
3. Laser steering unit works on it as the fast scan axis.
• Y-axis Motor Issue
1. It happens the regular horizontal lines.
2. It can't pull out the moving plate from the cassette housing when it starts to scan.
Exitation
Emission
IP
Fast scan axis
S
low
sca
n
a
x
is
Содержание FireCR
Страница 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Страница 5: ...for User manual download for Quantor software download...
Страница 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Страница 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Страница 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Страница 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Страница 20: ...Main Board Core Board...
Страница 23: ...BLDC board Function 1 BLDC motor driver 2 Control BLDC motor to maintain same stable speed BLDC Issue 1 Bad RPM...
Страница 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Страница 41: ...2 Disassembly Top Cover 1 Pull back cover...
Страница 47: ...2 Disassembly BLDC Board 1 Remove Top Cover 2 Remove Internal Cover 3 Unplug 2 wire connectors from BLDC Board...
Страница 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Страница 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Страница 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Страница 59: ...2 Disassembly 1 Remove the 6 screws for from the fiber bundle cover 2 Lift off cover Fiber Bundle...
Страница 60: ...2 Disassembly 3 Remove the two screws from the PMT housing 4 Loosen PMT connectors Fiber Bundle...
Страница 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...
Страница 63: ...Calibration and Images Note Exposure must always cover the entire cassette Calibration Geometry...
Страница 74: ......