Calibration and Images
Understanding the Cal-files:
Calibration files can be found in the software folder. Windows Photo Viewer
can view the Cal-
files. We recommend that you open the “High Dose”
calibration of each cassette size to verify an artifact free calibration.
14 x 17 Cassette Calibration
10 x 12 Cassette Calibration
File
Description
File
Description
Cal0.tif
Scan Blank
Cal4.tif
Scan Blank
Cal1.tif
Scan Low Dose
Cal5.tif
Scan Low Dose
Cal2.tif
Scan Mid Dose
Cal6.tif
Scan Mid Dose
Cal3.tif
Scan High Dose
Cal7.tif
Scan High Dose
Section.dat
Cal0 - Cal3
compressed together
Section2.dat
Cal4 - Cal7
compressed together
Содержание FireCR
Страница 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Страница 5: ...for User manual download for Quantor software download...
Страница 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Страница 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Страница 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Страница 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Страница 20: ...Main Board Core Board...
Страница 23: ...BLDC board Function 1 BLDC motor driver 2 Control BLDC motor to maintain same stable speed BLDC Issue 1 Bad RPM...
Страница 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Страница 41: ...2 Disassembly Top Cover 1 Pull back cover...
Страница 47: ...2 Disassembly BLDC Board 1 Remove Top Cover 2 Remove Internal Cover 3 Unplug 2 wire connectors from BLDC Board...
Страница 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Страница 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Страница 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Страница 59: ...2 Disassembly 1 Remove the 6 screws for from the fiber bundle cover 2 Lift off cover Fiber Bundle...
Страница 60: ...2 Disassembly 3 Remove the two screws from the PMT housing 4 Loosen PMT connectors Fiber Bundle...
Страница 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...
Страница 63: ...Calibration and Images Note Exposure must always cover the entire cassette Calibration Geometry...
Страница 74: ......