Common Issue
1. No Power
1.1 Check to see if power supply is in working order by swapping with a working one.
1.2 Remove top cover and internal cover
.
7.3 Check the voltage.
7.4 Make sure that voltage reads 24V (Scanner must turned on).
7.5 If not, disconnect the power cable on the main board
7.6 Make sure that voltage reads 24V
7.7 If not, replace power board or power switch.
- Refer to the Service Manual p.33 for power board replacement
Содержание FireCR
Страница 1: ...DIGIRAY CORP Basic Guide of FireCR Scanner...
Страница 5: ...for User manual download for Quantor software download...
Страница 10: ...the Windows Security Dialog Box pops up after installation click Allow access...
Страница 11: ...7 PC requirement Recommended PC requirement Minimum PC requirement...
Страница 13: ...Basic configuration of CR Computed Radiography system Imaging Plate Cassette Housing...
Страница 16: ...FireCR Scanner Video on how to use https www youtube com watch v xVkq_ sIabc...
Страница 20: ...Main Board Core Board...
Страница 23: ...BLDC board Function 1 BLDC motor driver 2 Control BLDC motor to maintain same stable speed BLDC Issue 1 Bad RPM...
Страница 33: ...Understanding FireCR Flash Calibration Note X ray exposure must cover entire cassette...
Страница 41: ...2 Disassembly Top Cover 1 Pull back cover...
Страница 47: ...2 Disassembly BLDC Board 1 Remove Top Cover 2 Remove Internal Cover 3 Unplug 2 wire connectors from BLDC Board...
Страница 48: ...2 Disassembly BLDC Board 4 Remove 4 screws 2 5mm hex 5 Reinstall in reverse order...
Страница 49: ...2 Disassembly Main Board 1 Unscrew the 6 screws that hold down the board and replace 2 5mm hex...
Страница 54: ...2 Disassembly Aligner Motor 1 Remover back over...
Страница 59: ...2 Disassembly 1 Remove the 6 screws for from the fiber bundle cover 2 Lift off cover Fiber Bundle...
Страница 60: ...2 Disassembly 3 Remove the two screws from the PMT housing 4 Loosen PMT connectors Fiber Bundle...
Страница 61: ...2 Disassembly 5 Remove the two screws at the edge of the eraser 6 Pull back eraer Fiber Bundle...
Страница 63: ...Calibration and Images Note Exposure must always cover the entire cassette Calibration Geometry...
Страница 74: ......