Installation Guide
Revision 2.00 of 04/11/2009
Page 9 of 33
5
.
P
LACEMENT
The environment intended for this instrument is the analysis laboratory.
For safety reasons and given the type of exams that it executes, the instrument has to be
placed far from sources of heat, in zones non accessible to liquids, in environments free from
dust and on perfectly flat work benches that are not subject to shocks or vibrations.
Furthermore it is advised that the VesMaticCube be placed far from possible generators of
Electro magnetic waves (for example fridges, laboratory centrifuges) and from instrumentation
without the CE mark, since they could affect the proper functioning of the instrument.
It is advised that a proper table be used that can support the weight of the instrument. The table
or bench top should not exceed 90cm in height, to guarantee an ergonomically correct
position for the operator during the input of the commands on the Tablet PC.
The table or bench top where the instrument will be placed, should allow enough space, about
40cm, on the sides of the instrument for the operator to easily introduce and extract the sample
holder rack in the classifier module (Fig. 5.1)
Fig. 5.1 ‘Frontal view with the extensions for rack input’
Furthermore, to be able to reach the connectors on the rear of the window and, most of all, to
be able to quickly access the switch and the power cable in case of emergency, it is necessary
to maintain a
For operator safety, do not place any materials or objects such as paper or containers on or
near the instrument.
safe distance from the wall of at least 20 cm from the back side of the instrument.
Choose a position close to an undisturbed outlet free from electrical fluctuations.