Installation Guide
Revision 2.00 of 04/11/2009
Page 3 of 33
1.
T
ECHNICAL
S
PECIFICATIONS
Power supply
Europe: 230Vac@50Hz; Usa/Canada: 110-120Vac@60Hz
Absorbed electrical power 265VA
Fuses
2 x 5,0A T (5 x 20 mm) UL-CSA
Dimensions
650 x 580 x 690 mm ( l x h x p )
Weight
45 Kg
Ambient temperature
Operational
from +15 to +35°C
Storage
from + 5°C to + 45°C
Relative humidity threshold from 20 to 80% without condensation
Central unit
Freescale i.MX31 ARM11 Microprocessor; Flash 128MB NAND;
128 MB DDR RAM
Display
TFT 800x 600 colours with Touch Screen
Peripheral control unit
Microprocessor board on host Bus
Internal analytical section Chain with 89 positions for the relative test tubes
Chain advancing step
19 seconds in the analytical cycle
Sample insertion section 10 + 10 sliding blocks for housing and transporting the typical
haematology racks
Analysed sample
collection section
Sample-holder racks with 8x14 positions for storing
processed test tubes.
Optical units
Two pairs of opto-electronic units (Led & Analogue Sensor).
Printer
Alphanumerical with thermal paper 58 mm wide, 36
characters per row, speed 20 mm/sec.
Interfaces
2 x RS232C, 2 USB Host, 1 USB Client, 1 Slot Compact Flash
Protection Category
CLASS I
Safety standards
CEI EN 61010-1 (Ed.2001-11); CAN/CSA-C22.2 No.61010-1-04
(Ed.2004-07); UL61010-1 (Ed.2004-07)
EMC
CEI EN 61326 (Ed.2004-08)
Installation category
II
The safety and performance standards of the instrument are not guaranteed in the
case of a different power supply cable from the one supplied being used with the
instrument
.