Installation Guide
32 of 33
Revision 2.00 of 04/11/2009
Case 2
0
20
40
60
80
100
120
140
0
20
40
60
80
100
120
140
run 1
run 2
20,62
In this case the test has to be considered not correct since the repetition of a sample changes
its state, from normal to pathologic and vv.
Note 1:
Before processing the samples assure that they are:
a)
fresh or stored at 4°c
b)
wait for them to arrive at room temperature
WARNING:
Avoid analyzing samples directly from the fridge, the subsequent warming up will cause the
test to be poorly reproducible for a progressive modification
in the behavior of the
sedimentation.
Note 2:
Before reporting the obtained values in a test of reproducibility pay attention of possible error
flags present. (see paragraph on errors)