7.2.7. Checking of the Power Supply
Open the instrument for regular cleaning inside. Check the cleanliness and operation of
the cooling fan of the power supply. Clean or replace the fan if necessary.
7.2.8. Bleaching of Fluidic System
It is recommended to run a bleaching procedure to remove stains from the fluidic
system.
1. Connect 2-5%, hand warm, clean bleach solution to all reagent inputs, and perform
priming on all reagent inputs.
2. Leave it in the tubing for not more than 2-3 minutes.
3. Remove the bleach, prime on air.
4. Connect distilled water (100 ml), and perform priming all reagents, again.
Connect reagents, and run priming again.