Abacus
4
Service Manual
50
Code
Error text
Reason(s)
Remedy
1107
Check Cleaner container!
Cleaner tank is
empty,
or
the
dil./cleaner sensor
is too sensitive (try
recalibrating
the
sensors), or there
are lot of bubbles
in the cleaner tube
Replace
the
cleaner
reagent
container with a filled one.
If this is still a problem with a filled
cleaner container, run Reagent
sensor calibration and check the
tubing and fittings of cleaner path.
1110
Check Rinse container!
Rinse
tank
is
empty, or the rinse
sensor
is
too
sensitive
(try
recalibrating
the
sensors), or there
are lot of bubbles
in the rinse tube
Replace the rinse reagent container
with a filled one.
If this is still a problem with a filled
rinse container, run Reagent sensor
calibration and check the tubing and
fittings of rinse path.
1111
The sampler rotor was jammed
during sampling process!
Retry sampling?
Cap was not taken
off
from
the
sample tube, or
tube is not placed
correctly, or other
mechanical jam
Check the adapter and the tube.
Press Retry or Stop.
1112
The
sampler
rotor
was
jammed!
Mechanical
problems like
adapter is moved
out from the right
place
Check the adapter and the tube.
Press the
button
1113
Unrecoverable
pneumatical
error occurred!\n
Device: %s\n
Error code: %s\n
Status: %x\n
Please turn off the instrument
and turn it on again!
Fatal error in
mechanical
subsystem.
Possibly
mechanical jam.
Check the proper operation of the
motor modules (home/end opto
switches, ribbon cables and the
motor moving).
1201
Fatal HGB error.
Power off the system!
HGB channel did
not give a ready
signal!
Replace the COMB board.
1400
Printer error %d
Printer error!
Check the connections between
instrument and printer and check
printer setup in Printer Settings