Abacus
4
Service Manual
41
5.2.5. Service Calibration
The analyzer provides a menu for Service calibration purposes.
In result calculations the service calibration factors are used as the user calibration factors, so
they are multiplied for each parameter:
RBC
Disp.
= Fact
RBC User
* Fact
RBC Serv.
* RBC
Measured
If the user factor is near the bound (0.80 - 1.20), by setting the corresponding service factor, the
user factor can be adjusted to 1.00.
Example:
Fact
RBC User
= 1.19 and
Fact
RBC Serv
= 0.96, and
Fact
RBC User
= 1.00 and
Fact
RBC Serv
= 1.14 gives the same result for RBC.
Apply user
calibration
factors
function is used to combine user and service calibration factors.
The software will multiply the existing factors, and move them to the Service level to set user
factors to 1.00.
5.2.5. Service Settings
Service Person can set the parameters of the automatic functions in this dialog menu.
Autoclean cycle
: after how many measurements the instrument makes an autocleaning
Language:
this setting will adjust the user
language. The Service menu always uses English
language.
Database capacity:
Can be set to 1000, 2000,
5000 records, default is 1000. Any change in the
database capacity will lead to loss of previous
records, the instrument warns the service engineer
prior to delete the data.
Laboratory header size
: you can choose how
many lines to contain the header of the blood
result.
Calendar mode:
You can select between the Gregorian and the Jalaali (Persian) calendar. All
dates will be converted accordingly.
Startup dialog:
you can choose the menu which
the instrument enters after power on
Disable multi-user mode:
by this setting the
multi-user mode can be disabled.
3 part mode:
by selecting this mode only the 3
part parameters are displayed, the EOS count is
not performed.
Reag pack mode:
this option must be selected if
the customer uses a reagent pack.