32
3.6. Lysing process
In this step the set lysing reagent is added into the measuring chamber through V9 (On), while
the Lyse syringe moves upwards. This process makes the WBC/HGB dilution with lyse reagent.
For better mixing the macro syringe pushes some air bubbles (aspirated through the washing
inlet of the chamber and V3) after the lysing process through V8 (Off), V7 (On), V4 (Off) V3
(On).