![Diamond SmartLyte Plus Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/diamond/smartlyte-plus/smartlyte-plus_service-manual_2493456018.webp)
1 Introduction
ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 18 of 58
Figure 1-16 Connectors and cables on CPU Board
3. On the CPU board, unclip the top and bottom tabs of the valve board connector. Holding the
locking tabs open at the same time helps to push the ribbon cable connector out of the CPU
connector.
4. Move the valve ribbon cable to the side.
5. Remove the 6 screws with their plastic washers that hold the CPU board to front case.
6. Carefully pull the CPU board off the alignment pin (
RFID board and the printer is not dislodged.
7. Handle board carefully so as to not damage the Electrode Pins on the front of the CPU
board.
8. Each Pin can be removed by pulling the Electrode Pin from its socket. (
Alignment Pin
Display Cable
Valve Board
Connector
Power Cable
Connector
Top Grounding
Screw
CPU Screws, 6
Interface PC
Ribbon Cable
Содержание SmartLyte Plus
Страница 1: ......