Document Title:
Remote Tomo5 operation guide for eBIC microscopes
Author(s):
Kyle Morris
Julika Radecke
Andrew Howe
Version:
2.0.3
Date:
27 Jun 2022
If you have any questions at any point during set-up, please ask your Local Contact for assistance Page 9 of 29
Atlas collection:
Collect an Atlas of the chosen grid(s)
Atlas > Screening
Select the checkbox of each grid you want to acquire an Atlas of and select Close Col Valves
-
If acquiring more than one Atlas, go to
autoloader and switch the turbo to “Always On”.
o
Wait for the Turbo to reach 100 %
o
Check Turbo status on the microscope PC
▪
TEM Life Science Autoloader Vacuum control
-
Switch turbo to “Auto Off” once you have finished taking Atlas maps of your grids.
-
Select “Close Col. Valves” if you leave the microscope unsupervised
Click the ‘Start’ button