Document Title:
Remote Tomo5 operation guide for eBIC microscopes
Author(s):
Kyle Morris
Julika Radecke
Andrew Howe
Version:
2.0.3
Date:
27 Jun 2022
If you have any questions at any point during set-up, please ask your Local Contact for assistance Page 10 of 29
3.1.3 Inspect your Atlas
Inspect your single or multiple Atlases for targets by clicking on the grid in the left panel
-
Left-click drag to move around, middle scroll to zoom in-out
When you have chosen your grid for target setup, select it and click ‘Load Sample’ in Tomo
3.1.4 Check your imaging condition Presets
Navigate to a square in your Atlas, right-click > Move stage here
In the Preparation tab > Acquisition and Optics Settings, you may select the Preset parameters
and adjust to suit your experiment.
Check your low magnification Presets using your loaded grid
Take a Preview using the Atlas Preset
-
If necessary, re-centre on the square
Take a Preview using the Overview Preset
-
If necessary, re-centre on the square and reacquire the Preview
-
Adjust the magnification to fill the field of view with the grid square
-
For lamella, the Overview preset should be adjusted to fill the field of view with one
lamella
Important: Only load grids from inside Tomo,
any shifts and rotations that occur to the grid during
loading and unloading will be measured and accounted for automatically. If you notice navigation on the
Atlas places you in the incorrect area this may have failed and you will need to reacquire your Atlas.
Overview - 540 X
Overview - 320
X