I 19 I
DIALAB PHOTOMETER DTN-410K USER MANUAL
Place a test tube with water in sipper and press pump actuator.
PUMP ACTUATOR
Repeat until absorbance readings are stable /- 0.005 Abs.
If readings are not stable, aspirate solution #1 and wait for 2 or 3 minutes. Then aspirate the
same solution again, aspirate soaking solution #2 several times (3 at least). Finally, aspirate
water and verify stability of reading.
Pump Actuator
In Wash method: runs the pump
In Run method: aspirates blank, standard and sample
Содержание DTN-410 K
Страница 1: ...I 1 I DIALAB PHOTOMETER DTN 410K USER MANUAL PHOTOMETER DTN 410 K USER MANUAL ...
Страница 2: ...I 2 I DIALAB PHOTOMETER DTN 410K USER MANUAL ...
Страница 6: ...I 6 I DIALAB PHOTOMETER DTN 410K USER MANUAL ...
Страница 14: ...I 14 I DIALAB PHOTOMETER DTN 410K USER MANUAL ...
Страница 20: ...I 20 I DIALAB PHOTOMETER DTN 410K USER MANUAL ...