DLPT2+
Operating Instructions
25
3.3
Testing a Class II Appliance
•
Visually inspect the appliance and mains cord
for signs of damage.
•
If the appliance passes a visual inspection
proceed with the electrical tests.
•
Plug the earth test lead into the Earth Continuity
Test Terminal on the DLPT2+ end panel.
•
Plug the appliance into the DLPT2+ front panel
test/mains outlet socket.
•
Connect the earth test probe to an exposed
metal part on the appliance.
•
If the Appliance under test has an ON/OFF
switch, make sure it is in the ON position.
•
Press the CLII test key
•
The display will momentarily show the insulation
test voltage before proceeding with the test.
Note: The power switch on the appliance under
test must be in the ON position to perform
an insulation test. If no appliance is
detected the DLPT2+ will display the
following.
Содержание DLPT2+
Страница 1: ......
Страница 9: ...DLPT2 Operating Instructions 9 Figure 1 DLPT2 Front View ...
Страница 10: ...DLPT2 Operating Instructions 10 Figure 2 DLPT2 End View Figure 3 Testing CLI appliances no mains leakage ...
Страница 11: ...DLPT2 Operating Instructions 11 Figure 4 Testing CLII appliances no mains leakage ...
Страница 12: ...DLPT2 Operating Instructions 12 Figure 5 Testing CLI appliances mains leakage ...
Страница 13: ...DLPT2 Operating Instructions 13 Figure 6 Testing CLII appliances mains leakage ...
Страница 14: ...DLPT2 Operating Instructions 14 Figure 7 Testing IEC Leads ...
Страница 15: ...DLPT2 Operating Instructions 15 Figure 8 Testing Installation RCD ...
Страница 16: ...DLPT2 Operating Instructions 16 Figure 9 Testing Plug Top RCD ...