background image

MPG2-H OPERATION AND MAINTENANCE MANUAL 

© 2004-2008 DH Instruments, a Fluke Company 

Page 10 

During operation, if the VV runs out of stroke, the test shut off and reservoir shutoff valves 
can be used in sequence to allow the variable volume to be refilled without releasing the 
pressure on the TEST1 and TEST2 port (see Section 3.2.3). 

 

  The maximum operating pressure of the variable volume is 200 MPa (30 000 psi) with 

MPG2-H and 100 MPa (15 000 psi) with MPG-H-HV and.  Do not generate pressure greater than 
the maximum operating pressure. 

3.1.5

 

RESERVOIR 

The reservoir is a 300 cc (18 in

3

) aluminum vessel mounted inside the MPG2 housing.  It is 

accessed for filling or cleaning by a screw on cover on the top surface of the MPG2 housing. The 
reservoir is connected to the RESERVOIR ISOLATION valve by flexible tubing. The reservoir lid 
can used to force fluid out of the reservoir into the system connected to the OPG1. 

3.2

 

OPERATION 

Operation of the MPG2 can be divided into three common tasks: filling and priming the system under test 
(See Section 3.2.1); generating and adjusting pressure (see Section 3.2.2); filling or emptying the variable 
volume while maintaining test system pressure (see Section 3.2.3). 

3.2.1

 

FILLING AND PRIMING THE SYSTEM UNDER TEST, 

PRIMING PUMP 

It is important to remove air from the test system by filling and priming it before using the 
variable volume to generate pressure.  Failure to fill and prime will result in the need to use 
multiple variable volume strokes to generate pressure. 

The reservoir pump is designed to push oil through the MPG2 and out the TEST port into the 
test system.  It can also generate an initial pressure, saving variable volume stroke. 

To fill and prime the test system, prior to generating pressure with the variable volume, 
proceed as follows: 

  Leave an open purge point (for example a loose fitting) at the highest point possible in 

the system that is connected to the MPG2. 

  Open the reservoir shut off valve and the test shut off valve to connect the reservoir to 

the rest of the system. 

  Position the reservoir cover correctly by rotating it CW until the reservoir vent valve is at 

the CLOSE position.   Do not over tighten; do not turn CW beyond CLOSE position. 

  Be sure that the reservoir cover vent valve is open (loosened CCW).   

 

Pull the reservoir handle pump up slowly to fill it with oil from the reservoir.  Then push 
the handle down firmly with moderate speed to seal the tip of the pump against the 
reservoir seat and inject oil into the MPG2 and out the TEST port. 

 

Repeat   while observing the open purge point in the test system.  When only oil and no 
air comes out of the purge point, close and tighten the purge point. 

 

If desired, repeat   to set an initial pressure with the reservoir pump. 
Be sure the variable volume is already filled and in the start position (Section 3.2.2,  - ), 
and the reservoir shut off valve is open. 

Содержание MPG2-H

Страница 1: ...2004 2008 DH Instruments a Fluke Company MPG2 H and MPG2 H HV Manual Pressure Generator Controller for Hydraulic Pressure Operation and Maintenance Manual...

Страница 2: ...H Instruments makes sincere efforts to ensure the accuracy and quality of its published materials however no warranty expressed or implied is provided DH Instruments disclaims any responsibility or li...

Страница 3: ...ON 7 2 2 SITE REQUIREMENTS 7 2 3 INITIAL SETUP 8 3 OPERATION 9 3 1 GENERAL OPERATING PRINCIPLES AND INFORMATION 9 3 1 1 OVERVIEW 9 3 1 2 VALVES 9 3 1 3 PRIMING PUMP 9 3 1 4 VARIABLE VOLUME VV 9 3 1 5...

Страница 4: ...TA AB BL LE ES S Table 1 MPG2 H and MPG2 H HV Parts List 7 Table 2 DHI Authorized Service Providers 16 F FI IG GU UR RE ES S Figure 1 Front view 2 Figure 2 Rear view 2 Figure 3 Top view 3 Figure 4 Fro...

Страница 5: ...f its many features and functions For those of you who don t read manuals go directly to Section 2 3 to set up your MPG2 Then go to Section 3 2 This will get you running quickly with minimal risk of c...

Страница 6: ...MPG2 H OPERATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...stroke needed to generate the desired pressure The maximum working pressure of MPG2 H HV is 100 MPa 15 000 psi The MPG2 H requires less variable volume torque and adjusts pressure more precisely 1 2 S...

Страница 8: ...ir cover vent valve 3 Priming pump 4 Reservoir cover 5 Reservoir shut off valve 6 Variable volume 7 Test shut off valve and fine pressure adjustment Figure 1 Front view 1 Reservoir 2 Test shut off val...

Страница 9: ...ny 1 TEST1 connection 2 TEST2 connection 3 Reservoir cover vent valve 4 Priming pump 5 Reservoir cover 6 Test shut off valve and fine pressure adjustment 7 Variable volume 8 Variable volume piston pos...

Страница 10: ...ATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page 4 1 4 DIMENSIONAL SCHEMATICS All dimensions are in millimeter mm Figure 4 Front view dimensions Figure 5 Bottom and side view...

Страница 11: ...Company 1 5 HYDRAULIC SCHEMATIC 1 Reservoir cover 2 Priming pump 3 Reservoir cover vent valve 4 TEST1 connection 5 TEST2 connection 6 Test shut off valve and fine pressure adjustment 7 Variable volum...

Страница 12: ...MPG2 H OPERATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page 6 N NO OT TE ES S...

Страница 13: ...strument 3323107 or 3341293 Accessories including 4 ea Variable volume handle 3329907 1 ea Plug DH500 3068470 1 Operation and Maintenance Manual 3340765 2 2 SITE REQUIREMENTS MPG2 should be installed...

Страница 14: ...connections are DH500 female DH500 is a gland and collar type fitting for 1 4 in 6 mm coned and left hand threaded tube DH500 is equivalent to AE F250C HIP HF4 etc If a TEST port is not being used pl...

Страница 15: ...valve needle displaces fluid when it is operated it can also be used as a very fine variable volume for very fine pressure adjustment 3 1 3 PRIMING PUMP The priming pump is a syringe style fluid inje...

Страница 16: ...ming it before using the variable volume to generate pressure Failure to fill and prime will result in the need to use multiple variable volume strokes to generate pressure The reservoir pump is desig...

Страница 17: ...the system to which it is connected should be as free of air as possible See Section 3 2 1 to fill and prime the test system Once the system to which MPG2 is connected filled and primed to generate an...

Страница 18: ...lows Close the test shut off valve This isolates the system connected to the TEST port s from the variable volume and the reservoir shut off valve Note that after refilling the variable volume the tes...

Страница 19: ...r Remove the balance of liquid from the system by screwing the variable volume all the way in CW with the TEST1 port plugged and the TEST2 port open Then use the filling and priming procedure describe...

Страница 20: ...MPG2 H OPERATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...

Страница 21: ...after receiving authorization from DHI or its Authorized Service Provider The buyer assumes all liability vis vis third parties in respect of its acts or omissions involving use of the products In no...

Страница 22: ...02 431 9559 cal repair dhinstruments com Ohte Giken Inc Technology Center 258 1 Nakadai Nihari Gun Ibaraki 300 0133 JAPAN Tel 0298 55 8778 Fax 0298 55 8700 tohte ohtegiken co jp DHI Products Technical...

Отзывы: