background image

3. OPERATION 

 

Page 9 

© 2004-2008 DH Instruments, a Fluke Company   

3

3

.

.

 

 

O

O

P

P

E

E

R

R

A

A

T

T

I

I

O

O

N

N

 

 

3.1

 

GENERAL OPERATING PRINCIPLES AND INFORMATION 

3.1.1

 

OVERVIEW 

See Section 1.5, Figure 6 

MPG2-H is used to manually generate and adjust liquid pressure from atmosphere to 
200 MPa (30 000 psi) (limited to 100 MPa (15 000 psi) with MPG2-H-HV) or using a priming 
pump, two valves and a variable volume.  Fittings for connecting to a device or system under 
test are located on the top and middle rear of the instrument housing.  A reservoir holds a 
supply of fluid. 

3.1.2

 

VALVES 

The reservoir shutoff and test shutoff valves are manually operated needle valves. 

The reservoir shut valve isolates the reservoir from the pressurized circuit. It is opened use 
the priming pump, to fill the variable volume with fluid and to open the pressurized circuit to 
atmosphere (vent). 

The test shutoff valve isolates the TEST1 and TEST2 ports from the rest of the MPG2 circuit.  
It can be used to isolate the variable volume and reservoir from the pressurized test system, 
for example, if it becomes necessary to refill the variable volume with fluid when it reaches 
end of stroke. 

As test shutoff the valve needle displaces fluid when it is operated, it can also be used as a 
very fine variable volume for very fine pressure adjustment. 

3.1.3

 

PRIMING PUMP 

The priming pump is a syringe style fluid injector, built into the reservoir cover.  On its plunger 
upstroke, it draws oil from the reservoir into its barrel.  On the down stoke, the barrel tip seals 
with the connection on the bottom of the reservoir and the plunger pushes oil into the rest of 
MPG2 and out the test port. 

The priming pump is used to fill the MPG2 and system to which it is connected with hydraulic 
fluid and then generate a low, starting pressure. 

3.1.4

 

VARIABLE VOLUME (VV) 

The variable volume (VV) is a piston-cylinder type screw press.  Rotating the variable volume 
handles, moves the piston in or out decreasing or increasing the volume and therefore 
increasing or decreasing pressure. 

The VV is used to generate and adjust pressure in the system to which MPG2 is connected 
when the reservoir shut valve is closed and test shut off valve is open.  The torque necessary 
to rotate the VV handle is proportional to the pressure in the variable volume. 

Generally, the starting position of the VV when beginning to generate pressure from 
atmosphere is with the piston fully retracted (handles rotated maximum CCW) so that the full 
stroke of the VV is available to fill the test system and generate pressure. 

Содержание MPG2-H

Страница 1: ...2004 2008 DH Instruments a Fluke Company MPG2 H and MPG2 H HV Manual Pressure Generator Controller for Hydraulic Pressure Operation and Maintenance Manual...

Страница 2: ...H Instruments makes sincere efforts to ensure the accuracy and quality of its published materials however no warranty expressed or implied is provided DH Instruments disclaims any responsibility or li...

Страница 3: ...ON 7 2 2 SITE REQUIREMENTS 7 2 3 INITIAL SETUP 8 3 OPERATION 9 3 1 GENERAL OPERATING PRINCIPLES AND INFORMATION 9 3 1 1 OVERVIEW 9 3 1 2 VALVES 9 3 1 3 PRIMING PUMP 9 3 1 4 VARIABLE VOLUME VV 9 3 1 5...

Страница 4: ...TA AB BL LE ES S Table 1 MPG2 H and MPG2 H HV Parts List 7 Table 2 DHI Authorized Service Providers 16 F FI IG GU UR RE ES S Figure 1 Front view 2 Figure 2 Rear view 2 Figure 3 Top view 3 Figure 4 Fro...

Страница 5: ...f its many features and functions For those of you who don t read manuals go directly to Section 2 3 to set up your MPG2 Then go to Section 3 2 This will get you running quickly with minimal risk of c...

Страница 6: ...MPG2 H OPERATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...stroke needed to generate the desired pressure The maximum working pressure of MPG2 H HV is 100 MPa 15 000 psi The MPG2 H requires less variable volume torque and adjusts pressure more precisely 1 2 S...

Страница 8: ...ir cover vent valve 3 Priming pump 4 Reservoir cover 5 Reservoir shut off valve 6 Variable volume 7 Test shut off valve and fine pressure adjustment Figure 1 Front view 1 Reservoir 2 Test shut off val...

Страница 9: ...ny 1 TEST1 connection 2 TEST2 connection 3 Reservoir cover vent valve 4 Priming pump 5 Reservoir cover 6 Test shut off valve and fine pressure adjustment 7 Variable volume 8 Variable volume piston pos...

Страница 10: ...ATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page 4 1 4 DIMENSIONAL SCHEMATICS All dimensions are in millimeter mm Figure 4 Front view dimensions Figure 5 Bottom and side view...

Страница 11: ...Company 1 5 HYDRAULIC SCHEMATIC 1 Reservoir cover 2 Priming pump 3 Reservoir cover vent valve 4 TEST1 connection 5 TEST2 connection 6 Test shut off valve and fine pressure adjustment 7 Variable volum...

Страница 12: ...MPG2 H OPERATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page 6 N NO OT TE ES S...

Страница 13: ...strument 3323107 or 3341293 Accessories including 4 ea Variable volume handle 3329907 1 ea Plug DH500 3068470 1 Operation and Maintenance Manual 3340765 2 2 SITE REQUIREMENTS MPG2 should be installed...

Страница 14: ...connections are DH500 female DH500 is a gland and collar type fitting for 1 4 in 6 mm coned and left hand threaded tube DH500 is equivalent to AE F250C HIP HF4 etc If a TEST port is not being used pl...

Страница 15: ...valve needle displaces fluid when it is operated it can also be used as a very fine variable volume for very fine pressure adjustment 3 1 3 PRIMING PUMP The priming pump is a syringe style fluid inje...

Страница 16: ...ming it before using the variable volume to generate pressure Failure to fill and prime will result in the need to use multiple variable volume strokes to generate pressure The reservoir pump is desig...

Страница 17: ...the system to which it is connected should be as free of air as possible See Section 3 2 1 to fill and prime the test system Once the system to which MPG2 is connected filled and primed to generate an...

Страница 18: ...lows Close the test shut off valve This isolates the system connected to the TEST port s from the variable volume and the reservoir shut off valve Note that after refilling the variable volume the tes...

Страница 19: ...r Remove the balance of liquid from the system by screwing the variable volume all the way in CW with the TEST1 port plugged and the TEST2 port open Then use the filling and priming procedure describe...

Страница 20: ...MPG2 H OPERATION AND MAINTENANCE MANUAL 2004 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...

Страница 21: ...after receiving authorization from DHI or its Authorized Service Provider The buyer assumes all liability vis vis third parties in respect of its acts or omissions involving use of the products In no...

Страница 22: ...02 431 9559 cal repair dhinstruments com Ohte Giken Inc Technology Center 258 1 Nakadai Nihari Gun Ibaraki 300 0133 JAPAN Tel 0298 55 8778 Fax 0298 55 8700 tohte ohtegiken co jp DHI Products Technical...

Отзывы: