2. INSTALLATION
Page 27
© 2004-2007 DH Instruments, a Fluke Company
Connect DUT to molbloc-S Outlet:
Since there are so many types and
sizes of DUTs that can be calibrated by the molbloc-S, and since there is no
standardized dimension for the DUT’s plumbing centerline to its base, the
molstic-S was not designed to accommodate the DUT directly on its platform.
For the sake of flexibility, the molstic-S is supplied with 50 cm of PFA tubing
to help make the inlet and outlet connections. Use of the adjustable DUT
stand (P/N 401934) will also facilitate the alignment of the centerline of the
DUT’s inlet plumbing connection with that of the molstic-S plumbing.
For best results, follow the DUT manufacturer’s recommendation for the
length and size of upstream and downstream straight plumbing.