2.2

 

SITE REQUIREMENTS 

The molstic-S should be installed on a stable work surface at a convenient height.  The bottom edge of 
the molstic-S platform has a rubberized material that will protect the work surface from marring and 
scratching.  This will also prevent the molstic-S from slipping on smooth surfaces. 

When installing the molstic-S, consideration should be given to where the gas supply to molbloc-S will be 
located.  Minimize the length of supply plumbing and keep it organized so that it does not cross the area 
in which the operator will be working. 

2.3

 

INITIAL SETUP 

2.3.1

 

PREPARATION FOR OPERATION 

 

Remove the plastic protective caps from the supply and outlet ports of the molstic-S. 

 

Determine whether the DUT will be installed upstream or downstream of the molstic-S. 

  Make the inlet gas supply and DUT connections per the instructions below. 

 

 

DUTs that have built in flow control capability, such as MFCs, (and are to be tested in 
flow control mode) must be tested in the upstream location.  Install this type of DUT 

upstream of the molbloc-S only.

 

 

2.3.2

 

INSTALLATION AND SWAPPING OF MOLBLOC-S 

Typically the molstic-S mounting system configured with 1/4 in. plumbing is used to accommodate 
the molbloc-S elements with 48 mm square bodies and 1/4 in. male VCR

®

 inlet fittings, 

designated 1E2-S and smaller.  In some cases, when running “low” flows in sub-atmospheric mode, 
it might be desirable to use molbloc-S elements that have 1/2 in. VCR inlet fittings on the 1/4 in. 
molstic-S mounting system. Use the inlet flange conversion kit, P/N 401935. 

The molstic-S mounting system configured with 1/2 in. plumbing can accommodate any 
molbloc-S elements with 48 mm square bodies, up to 2E3-S.  In order to be used on this molstic, 
any molbloc-S element designated 1E2-S and smaller must have its 1/4 in. VCR inlet fitting 
replaced with the 1/2 in. VCR end flange through the use of conversion kit P/N 401936. 

2.3.2.1

 

REPLACING VCR INLET FLANGES ON MOLBLOC-S 
(AS REQUIRED) 

Use the following procedure to replace the VCR inlet flange on the molbloc-S 
elements, using conversion kit P/N 401935 or 401936. 

Содержание MOLSTIC-S

Страница 1: ...2004 2007 DH Instruments a Fluke Company molstic S Engineered Mounting Systems for molbloc S Mass Flow Elements Operation and Maintenance Manual...

Страница 2: ...warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual or produ...

Страница 3: ...STIC S DUAL 1 4 IN WITH LOW OR MID FLOW METERING VALVE AND REGULATION 13 2 INSTALLATION 15 2 1 UNPACKING AND INSPECTION 15 2 1 1 REMOVING FROM PACKAGING 15 2 1 2 INSPECTING CONTENTS 15 2 1 2 1 ACCESSO...

Страница 4: ...EGULATOR MAXIMUM PRESSURE ADJUSTMENT 31 3 2 3 LOCKING THE BACK PRESSURE REGULATOR ADJUSTMENT OPTIONAL 31 3 2 4 BACK PRESSURE REGULATOR OPERATION 31 3 3 LEAK CHECK RECOMMENDATIONS 32 3 4 MOLSTIC S FLOW...

Страница 5: ...ure 7 molstic S Dual 1 4 in with Regulation 8 Figure 8 molstic S Dual 1 2 in with Regulation 9 Figure 9 molstic S Single 1 4 in with Low or Mid Flow Metering Valve 10 Figure 10 molstic S Dual 1 4 in w...

Страница 6: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page IV N NO OT TE ES S...

Страница 7: ...rovide pressure control capability to the user Calibration test devices that are capable of controlling flow must be installed upstream of the molbloc S if they are to be tested in control mode Use th...

Страница 8: ...ORDERING INFO DESCRIPTION P N FAM0009 molstic S molstic for molbloc S Option 01 1 Single Channel 1 4 in Plumbing Figure 1 molstic S Single 1 4 in Standard Configuration 1 Inlet adaptor 1 4 in tube x...

Страница 9: ...09 molstic S molstic for molbloc S Option 01 2 Single Channel 1 2 in Plumbing Figure 2 molstic S Single 1 2 in Standard Configuration 1 Inlet adaptor 1 2 in tube x 1 2 in male VCR 2 Flow shut off mete...

Страница 10: ...molstic S molstic for molbloc S Option 01 3 Dual Channel 1 4 in Plumbing Figure 3 molstic S Dual 1 4 in Standard Configuration 1 Inlet adaptor 1 4 in tube x 1 4 in male VCR 2 Particulate filter 2 mic...

Страница 11: ...olstic S molstic for molbloc S Option 01 4 Dual Channel 1 2 in Plumbing Figure 4 molstic S Dual 1 2 in Standard Configuration 1 Inlet adaptor 1 2 in tube x 1 2 in male VCR 2 Flow shut off flow meterin...

Страница 12: ...3 2 With 1 4 in Pressure Reducing Regulator 03 4 With 1 4 in Back Pressure Regulator Figure 5 molstic S Single 1 4 in with Regulation 1 Inlet quick connect includes stem with 1 4 in SWG not shown 2 Pr...

Страница 13: ...03 3 With 1 2 in Pressure Reducing Regulator 03 5 With 1 2 in Back Pressure Regulator Figure 6 molstic S Single 1 2 in with Regulation 1 Inlet quick connect includes stem with 1 2 in SWG not shown 2 P...

Страница 14: ...With 1 4 in Pressure Reducing Regulator 03 4 With 1 4 in Back Pressure Regulator Figure 7 molstic S Dual 1 4 in with Regulation 1 Inlet quick connect includes stem with 1 4 in SWG not shown 2 Pressur...

Страница 15: ...3 With 1 2 in Pressure Reducing Regulator 03 5 With 1 2 in Back Pressure Regulator Figure 8 molstic S Dual 1 2 in with Regulation 1 Inlet quick connect includes stem with 1 2 in SWG not shown 2 Pressu...

Страница 16: ...4 in Plumbing 04 2 With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 9 molstic S Single 1 4 in with Low or Mid Flow Metering Valve 1 Inlet adaptor 1 4 in tube...

Страница 17: ...With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 10 molstic S Dual 1 4 in with Low or Mid Flow Metering Valve 1 Inlet adaptor 1 4 in tube x 1 4 in male VCR 2...

Страница 18: ...Pressure Regulator 04 2 With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 11 molstic S Single 1 4 in with Low or Mid Flow Metering Valve and Regulation 1 Inlet...

Страница 19: ...r 04 2 With 1 4 in Low Flow Metering Valve Option 04 3 With 1 4 in Mid Flow Metering Valve Figure 12 molstic S Dual 1 4 in with Low or Mid Flow Metering Valve and Regulation 1 Inlet quick connect incl...

Страница 20: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...

Страница 21: ...2 1 ACCESSORIES INCLUDED WITH MOLSTIC S SHIPMENTS Table 1 Accessory Kit 1 4 in molstic S versions P N 401938 DESCRIPTION PART 1 molstic S Operation and Maintenance Manual 550130 1 16 mm ISO KF Blanki...

Страница 22: ...ility such as MFCs and are to be tested in flow control mode must be tested in the upstream location Install this type of DUT upstream of the molbloc S only 2 3 2 INSTALLATION AND SWAPPING OF MOLBLOC...

Страница 23: ...22 for use with the new flange Be careful not to damage this seal in any way as a leak can result which will cause errors during flow measurement Position the replacement VCR inlet flange on the end o...

Страница 24: ...sconnecting molbloc S elements with VCR connections Connecting Disconnecting and Mounting molbloc S with VCR Face Seal Fittings molbloc S Inlet Fitting In making the molbloc S inlet flow connection al...

Страница 25: ...cked off Never use wrenches to hold the molbloc S body 2 3 2 2 2 molbloc S Connections to molbox For the pressure connections use the molbox to molbloc S pressure tubes P N 401125 supplied with the mo...

Страница 26: ...n the DUT and the molbloc The gas supply to the molstic S has the following requirements The gas supply must be clean and free of particulate contamination in order to avoid damage to the molbloc For...

Страница 27: ...ade kits that are available in 1 4 in or 1 2 in P N 401880 and 401881 respectively Figure 16 molstic S with Supplied Regulator DHI also offers the supply only platform that includes the regulator with...

Страница 28: ...tem leak testing The pressure reducing regulators supplied by DHI are selected for the following characteristics High quality High sensitivity and stability Outlet control pressure range 0 to 100 psig...

Страница 29: ...om particulate contamination Consult factory for availability of 1 2 in inline filter kit Connect DUT to molstic S inlet Since there are so many types and sizes DUTs that can be calibrated by the molb...

Страница 30: ...TYLE VACUUM FLANGE FITTINGS The outlet connection system on the molbloc S body is the ISO KF style vacuum flange It utilizes an internal centering ring an external overpressure ring and clamp The seal...

Страница 31: ...e separated Remove the clamp centering ring and overpressure ring Never use wrenches to hold the molbloc S body or tighten the wing nut 2 5 2 1 MOLSTIC S USE WITH DUT UPSTREAM During operation the mol...

Страница 32: ...he Downstream Tee Assembly end fitting type is the 25 mm ISO KF style vacuum flange These fittings can be converted to either Swagelok tube fitting or VCR Face Seal fittings using one of the following...

Страница 33: ...as not designed to accommodate the DUT directly on its platform For the sake of flexibility the molstic S is supplied with 50 cm of PFA tubing to help make the inlet and outlet connections Use of the...

Страница 34: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 28 N NO OT TE ES S...

Страница 35: ...er to obtain the most accurate and stable set point 3 1 1 SYSTEM PRESSURIZAION Begin with system pressure vented Make sure that the system pressure is vented Set inlet pressure to minimum Turn the reg...

Страница 36: ...sure Vent the system pressure Use the procedure in Section 3 1 to re pressurize and verify the regulator pressure setting Readjust if necessary Replace regulator knob cover Place the regulator adjustm...

Страница 37: ...te the back pressure regulator adjustment knob clockwise in order to raise the pressure level upstream of the regulator Rotate the back pressure regulator adjustment knob counterclockwise in order to...

Страница 38: ...S and will provide the control of inlet pressure to the molbloc S If the DUT does not have its own flow control capability the inlet pressure to the molbloc S must be controlled by either an upstream...

Страница 39: ...the test system Adjust system flow rate Rotate the flow control valve adjustment knob to set the desired flow rate on the molbloc S Fine adjustments in flow rate can be achieved by combining adjustme...

Страница 40: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 34 N NO OT TE ES S...

Страница 41: ...com 4 2 UPGRADE OF MOLSTIC S WITH 1 4 IN METERING VALVE The 1 4 in molstic S is delivered with integrated shut off metering valves These valves provide a measure of flow control when calibrating devi...

Страница 42: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 36 N NO OT TE ES S...

Страница 43: ...stic S filter is a sealed 2 micron pleated mesh filter Swagelok SS 4FW VCR 2 It can be cleaned by backflushing with clean gas but must be removed from the molstic S When removing cleaning and re insta...

Страница 44: ...MOLSTIC S OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 38 N NO OT TE ES S...

Отзывы: