Mode / Model:
B
-
Mode: 12.5 MHz (B
-
Scan)
DGH1912
B
-
Mode: 20 MHz (B
-
Scan)
DGH1920
A
-
Mode: 10 MHz (A
-
Scan)
DGH6006
MI
(Unitless)
I
SPTA.3
(mW/cm
2
)
MI
(Unitless)
I
SPTA.3
(mW/cm
2
)
MI
(Unitless)
I
SPTA.3
(mW/cm
2
)
Sample Size
3.00
3.00
3.00
3.00
5.00
5.00
K
4.258
4.258
4.258
4.258
2.742
2.742
Mean (x)
0.149
0.105
0.169
0.155
0.133
0.522
StdDev (Sx)
0.0154
0.0260
5.20E
-
3
7.24E
-
3
3.24E
-
3
0.0258
Limit (X)
0.215
0.216
0.191
0.185
0.142
0.593
Mode / Model:
B
-
Mode: 35 MHz (UBM)
DGH1500
-
35
B
-
Mode: 50 MHz (UBM)
DGH1500
-
50
MI
(Unitless)
I
SPTA.3
(mW/cm
2
)
MI
(Unitless)
I
SPTA.3
(mW/cm
2
)
Sample Size
3.00
3.00
3.00
3.00
K
4.258
4.258
4.258
4.258
Mean (x)
0.177
0.0799
0.113
0.0262
StdDev (Sx)
3.61E
-
3
0.0116
8.72E
-
3
3.50E
-
3
Limit (X)
0.192
0.129
0.150
0.0411
ULTRASONIC EMMISSIONS INFORMATION
The following Acoustic Output Reporting Tables show the average outputs of the samples
tested. These results include the pressure
-
time waveform, the Fourier transform, the pulse in-
tensity integral and the waveform analysis taken at the derated spatial peak location (z
mjPII.3
).
The Statistical Analysis of Measurement Data tables show the results of a statistical analysis was
performed on the data to examine the upper output limits, based on a one
-
sided tolerance limit
approach. The mean and standard deviation of the Spatial
-
Peak, Time
-
Average Intensity and
Mechanical Index were found, and the upper output limits were calculated from the following
formula:
_____
X = x + K*Sx
__
where X is the upper output parameter limit, x is the average of the measured output
parameter, Sx is the standard deviation of the measured output parameter, and K is a factor
from Reference 11, Table A
-
7 (of
M.G.Natrella, Experimental Statistics, NBS Handbook 91, Octo-
ber 1963
). A value of K was chosen which corresponds to a 90% probability that 90% of all
probes would fall below the calculated limits X. Table notation definitions listed below:
MI
Mechanical Index
Z at max. I
pi.
α
derated spatial peak location
TIS
Soft Tissue Thermal Index
I
spta.3
derated spatial
-
peak pulse
-
average intensity
TIB
Bone Thermal Index
A
aprt
Active aperture area
TIC
Cranial Bone Thermal Index
F
c
center frequency (in MHz).
P
output power (in mW)
p
r
peak rarefactional pressure (in MPa)
T
d
pulse duration
F
awf
acoustic working frequenc (in MHz)
I
pi.
α
derated pulse intensity integral
STATISTICAL ANALYSIS OF MEASURMENT DATA