DS-CAL
SOFTWARE USER MANUAL
●
By checking “Generate Excel reports” Excel calibration certificates will be automatically
generated at the end of calibration procedure.
●
Leave the “Enable uploading reports to global database” unchecked. This is intended for internal
use only and is prepared for an automated upload of the calibration data to the DEWESoft server.
7.5. Tolerance
●
The warning limit defines the maximum allowable deviation of the measured value. By default it
is set to 70 % which means if a channel has an error higher than 70 % of the allowed specification
the channel will be marked in the overview in the color yellow during the calibration process. This
function is used for defining if an adjustment is necessary or not.
●
Leave the Tolerance mode at the default setting of “Default”; this means that the error
calculation is done based on the specified accuracy of the system. The setting “Production
tolerances” is used internally for functionality tests and has very tight limits.
Hint
The warning limit is very useful to diagnose if an adjustment should be done; the default value
of 70 % was determined by experience and makes sure that the unit will not be out of tolerance
until the next calibration after a year.
Illustration 15: Custom fields tab of the DSCalibrator
DS-CAL V20-1
24/67
Содержание DS-CAL
Страница 1: ...DS CAL SOFTWARE USER MANUAL DS CAL V20 1 1...
Страница 20: ...DS CAL SOFTWARE USER MANUAL Illustration 11 Users tab of the DSCalibrator DS CAL V20 1 20 67...
Страница 26: ...DS CAL SOFTWARE USER MANUAL Illustration 17 Prepare adjustment procedure window DS CAL V20 1 26 67...
Страница 55: ...DS CAL SOFTWARE USER MANUAL Illustration 46 Example of connecting the DSI 5A DS CAL V20 1 55 67...