DS-CAL
SOFTWARE USER MANUAL
4.3.2. FLUKE 55XX setup
The example below is based on FLUKE 5502E, but should apply in similar ways on the FLUKE 55XX series
of multifunction calibrators.
Go to Setup → Instmt Setup → Remote setup.
●
Host → serial
Go to Host setup where the following settings must be set:
●
8 data bits
●
1 stop bit
●
Stall → Xon/off
●
Parity → None
●
9600 baud
●
Remote I/F → Term
●
EOL → CRLF
Save the configuration by exiting the menus:
(Prev Menu) → (Prev Menu) → (Prev Menu) → Store changes? → Select Yes → (Prev Menu).
DS-CAL V20-1