8. Precautions

8.1) Be aware for any of the following conditions: 

There are additional conditions to take into account before using 

Negative  

Pressure Wound Therapy

, such as:

1.      

BLEEDING:

 There is a risk of bleeding/hemorrhaging with negative  

        

 pressure wound therapy. If hemostasis cannot be achieved, if the patient 
is on anticoagulants or platelet aggregation factors, or if the patient has 
friable blood vessels or infected vascular anastomosis, he or she may 
have an increased risk of bleeding; accordingly these patients should be 
treated in an inpatient care facility per their treating physician.

         

If active bleeding develops suddenly or in large amounts during  
therapy, immediately disconnect the pump, leave the dressing in place, 
and take measures to stop bleeding. Seek medical attention immediately.

2.      

VESSEL AND BONE PROTECTION:

 Precautionary measures should be  

        

   taken if any bones, vessels, ligaments or tendons are exposed. 

        Additionally, sharp edges (due to bone fragments) require special
        attention; these areas should be covered and smoothed wherever 
        possible. These conditions should be factored into the therapy 
        prescription as the attending clinician sees fit.

3.      

ENVIRONMENT:

         a. Defibrillation:  Remove the 

extriCARE™

 

dressing if defibrillation is  

 

required in the area of dressing placement.  Failure to remove the  
dressing may inhibit transmission of electrical energy and/or patient 
resuscitation.

         b. Magnetic Resonance Imaging (MRI):  The 

extriCARE™

 device is unsafe 

in the MR environment.  Do not take the 

extriCARE™

 

device into the MR 

environment. 

extriCARE™

 dressings however can typically stay on the 

patient with minimal risk in an MR environment, assuming that the use of 
the 

extriCARE™

 

Negative Pressure Wound Therapy System

 is not interrupted for 

more than two hours.

         c. Hyperbaric Oxygen Therapy (HBO):  Do not take the

 

extriCARE™

 

device 

into a hyperbaric oxygen chamber.  

extriCARE™

 devices are not designed 

for this environment, and should be considered a fire hazard in such an 
environment.  After disconnecting the 

extriCARE™

 device, either (i) replace 

the

 

extriCARE™

 

dressing with another HBO compatible material during the 

hyperbaric treatment, or (ii) cover the unclamped end of the 

extriCARE™

 

tubing. For HBO therapy, the 

extriCARE™

 tubing must not be clamped.  

Never leave an

 

extriCARE™

 

dressing in place without active 

extriCARE™

 

Negative Pressure Wound Therapy

 

for more than two hours.

Содержание extriCARE 2400

Страница 1: ...www devonmedicalproducts com 866 446 0092 Negative Pressure Wound Therapy System extriCARE 2400 Operating Manual IFU30 0003 Rev A 20121024...

Страница 2: ...onment in either intermittent or continuous mode This allows the user to program the specific pressure ranging from 40mmHg to 140mmHg In continuous mode the pressure is applied to the wound as long as...

Страница 3: ...or use Single Use Only Date Of Manufacture Type B Applied Part Internally powered electrical device Keep Dry Serial Number Prescription Use Only Power Switch Manufacture Lot Number Biohazard 2 SN LOT...

Страница 4: ...0 240 V 50 60 Hz 0 4A Output 5V 1 5A VACUUM MODES Continuous or Intermittent OPERATING CONDITIONS Temperature 5 C to 40 C 41 F to 104 F Humidity 15 to 85 non condensing PRESSURE OPTIONS 40mmHg 140mmHg...

Страница 5: ...ubing set EC2400 SO extriCARE NPWT Dressing Small oval with tubing set EC2400 S extriCARE NPWT Dressing Sacrum with tubing set EC2400 IM extriCARE NPWT Dressing Incision Management with tubing set EC2...

Страница 6: ...must be applied more frequently than the dressing change schedule allows 7 Warnings Review this manual prior to using the extriCARE 2400 Negative Pressure Wound Therapy Pump System If clarification is...

Страница 7: ...e factored into the therapy prescription as the attending clinician sees fit 3 ENVIRONMENT a Defibrillation Remove the extriCARE dressing if defibrillation is required in the area of dressing placemen...

Страница 8: ...sympathetic nervous system stimulation discontinue extriCARE therapy to minimize sensory stimulation and give immediate medical assistance 7 MODE In unstable anatomical structures continuous rather th...

Страница 9: ...ed 8 4 Dressing Management The extriCARE dressing is a one piece all inclusive dressing and should be removed in one piece In the event that the dressing comes apart all dressing materials must be rem...

Страница 10: ...he canister to the NPWT device 5 MODE SYMBOL Indicates pump operating mode continuous or intermittent 6 MODE BUTTON Allows user to set the pump to either continuous or intermittent mode 7 POWER PLUG E...

Страница 11: ...iCARETM System being used with an adapter that is not recommended There is a risk of volt age incompatibility if the input voltage is greater than 6V Less than 25 power remaining indicating that extri...

Страница 12: ...flashing on LCD screen Yellow LED flashing every 2 seconds Alarm symbol and 000 flashing on LCD screen Yellow LED flashing every 2 seconds Alarm symbol flashing on LCD screen Yellow LED flashing ever...

Страница 13: ...sing The dressing should be protected from moisture during bathing or changed prior to reconnecting to the pump Do not use the extriCARE 2400 Negative Pressure Wound Therapy Pump while show ering or b...

Страница 14: ...ft dressing material may be applied between each toe When used on the foot aggressive measures should be taken to protect the foot and divert unnecessary pressure If the dressing is applied over a new...

Страница 15: ...MODE Hold down MODE button for two seconds to select the mode continuous or intermittent A dotted line at the top right of the LCD Screen indicates intermittent treatment while a straight line indica...

Страница 16: ...ce will void the warranty In the event the extriCARE device requires repairs it should be returned to the medical equipment company or to Devon Medical Products directly Power adapter The extriCARE de...

Страница 17: ...en damaged in transit or has been subject to misuse neglect or accident or has been used in violation of Devon Medical Products instructions including without limitation the instructions contained in...

Страница 18: ...io frequency transmitting equipment and other sources of electrical noise in a healthcare environ ment high levels of interference may induce an abnormal stoppage or other disruption of this device Th...

Страница 19: ...for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5s 3 A m 6 kV contact 8 kV air 2 kV for power supply lines 1 kV line s and neutral 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in U...

Страница 20: ...higher the frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Fie...

Страница 21: ...ance according to frequency of transmitter m output power of transmitter W 0 01 0 1 1 10 100 150 kHz to 80 MHz d 1 2 P 0 12 0 38 1 2 3 8 12 80 MHz to 800 MHz d 1 2 P 0 12 0 38 1 2 3 8 12 800 MHz to 2...

Страница 22: ...50 69 Borquist O Ingemansson R et al Individualizing the Use of Negative Pre sure Wound Therapy for Optimal Wound Healing A focused Review of the Literature Ostomy Wound Management 2011 57 4 44 54 De...

Страница 23: ...Guidelines for Managing Pressure Ulcers with Negative Pressure Wound Therapy Advances in Skin and Wound Care 2004 17 1 16 Kirby M Negative Pressure Wound Therapy British Journal of Diabetes Vascular D...

Страница 24: ...www devonmedicalproducts com 866 446 0092...

Отзывы: