6. Contraindications for Use

The

 extriCARE™ 2400 System

 should 

NOT

 be used in the following conditions:

       

 •

  Exposed vessels, organs, or nerves.

       

 •

  Anastomotic sites.

     

   • 

  Exposed arteries or veins in a wound. All exposed vessels and organs in  

           

and around the wound must be completely covered prior to initiation of NPWT. 

Note:

 A thick layer of natural tissue is preferred.  Several layers of fine meshed 

non-adherent material or bio-engineered tissue may be an alternative. Ensure 
that protective materials will maintain their position throughout therapy.

        

  Fistulas, unexplored or non-enteric.

       

 •

  Untreated osteomyelitis.

       

 •

  Malignancy in the wound. The system may be used after excision of  

           malignancy if all margins are clear.
       

 •  

Excess amount of necrotic tissue with eschar. Note: The system may be  

           used with a minimal amount of slough or after debridement of necrotic tissue.
        

  Wounds which are too large or too deep to be accommodated by the dressing.

        

   Inability to be followed by a medical professional or to keep scheduled  

appointments.

       

 •

  Allergy to urethane dressings and adhesives. 

       

 • 

 Use of topical products which must be applied more frequently than the  

           dressing change schedule allows.

7. Warnings

        

   Review this manual prior to using the 

extriCARE™ 2400 Negative Pressure Wound 

Therapy Pump System

. If clarification is needed, contact technical personnel or 

Devon Medical Products at 866-446-0092 prior to use. Additional questions can be 
immediately addressed as well.   

        

   Do not use the 

extriCARE™ 2400 Negative Pressure Wound Therapy Pump

 around  

explosive or flammable material.  Do not use the pump in an MRI environment  
or hyperbaric chamber.  Disconnect prior to defibrillation.

       

 •

   This device should be used only under the direction of a trained professional,  

such as a doctor or nurse.

        

   Larger canister sizes (400cc or larger) should only be used in a facility where 

drainage can be closely monitored.

         •  

Negative Pressure Wound Therapy

 has not been cleared for use on children.

        •  

Use a properly rated charger to charge the lithium battery.  Incorrect     

           voltage and/or current can cause fire.

        •

  Do not place this device at temperatures greater than 170°F for more  

           than 2 hours which may cause a battery fire.

         •

   If battery swells up, gets hot, or smokes while charging, disconnect the charger 

immediately.  This may cause the battery to leak, and the reaction with air may 
cause the chemicals to ignite, resulting in fire.

5

6

Содержание extriCARE 2400

Страница 1: ...www devonmedicalproducts com 866 446 0092 Negative Pressure Wound Therapy System extriCARE 2400 Operating Manual IFU30 0003 Rev A 20121024...

Страница 2: ...onment in either intermittent or continuous mode This allows the user to program the specific pressure ranging from 40mmHg to 140mmHg In continuous mode the pressure is applied to the wound as long as...

Страница 3: ...or use Single Use Only Date Of Manufacture Type B Applied Part Internally powered electrical device Keep Dry Serial Number Prescription Use Only Power Switch Manufacture Lot Number Biohazard 2 SN LOT...

Страница 4: ...0 240 V 50 60 Hz 0 4A Output 5V 1 5A VACUUM MODES Continuous or Intermittent OPERATING CONDITIONS Temperature 5 C to 40 C 41 F to 104 F Humidity 15 to 85 non condensing PRESSURE OPTIONS 40mmHg 140mmHg...

Страница 5: ...ubing set EC2400 SO extriCARE NPWT Dressing Small oval with tubing set EC2400 S extriCARE NPWT Dressing Sacrum with tubing set EC2400 IM extriCARE NPWT Dressing Incision Management with tubing set EC2...

Страница 6: ...must be applied more frequently than the dressing change schedule allows 7 Warnings Review this manual prior to using the extriCARE 2400 Negative Pressure Wound Therapy Pump System If clarification is...

Страница 7: ...e factored into the therapy prescription as the attending clinician sees fit 3 ENVIRONMENT a Defibrillation Remove the extriCARE dressing if defibrillation is required in the area of dressing placemen...

Страница 8: ...sympathetic nervous system stimulation discontinue extriCARE therapy to minimize sensory stimulation and give immediate medical assistance 7 MODE In unstable anatomical structures continuous rather th...

Страница 9: ...ed 8 4 Dressing Management The extriCARE dressing is a one piece all inclusive dressing and should be removed in one piece In the event that the dressing comes apart all dressing materials must be rem...

Страница 10: ...he canister to the NPWT device 5 MODE SYMBOL Indicates pump operating mode continuous or intermittent 6 MODE BUTTON Allows user to set the pump to either continuous or intermittent mode 7 POWER PLUG E...

Страница 11: ...iCARETM System being used with an adapter that is not recommended There is a risk of volt age incompatibility if the input voltage is greater than 6V Less than 25 power remaining indicating that extri...

Страница 12: ...flashing on LCD screen Yellow LED flashing every 2 seconds Alarm symbol and 000 flashing on LCD screen Yellow LED flashing every 2 seconds Alarm symbol flashing on LCD screen Yellow LED flashing ever...

Страница 13: ...sing The dressing should be protected from moisture during bathing or changed prior to reconnecting to the pump Do not use the extriCARE 2400 Negative Pressure Wound Therapy Pump while show ering or b...

Страница 14: ...ft dressing material may be applied between each toe When used on the foot aggressive measures should be taken to protect the foot and divert unnecessary pressure If the dressing is applied over a new...

Страница 15: ...MODE Hold down MODE button for two seconds to select the mode continuous or intermittent A dotted line at the top right of the LCD Screen indicates intermittent treatment while a straight line indica...

Страница 16: ...ce will void the warranty In the event the extriCARE device requires repairs it should be returned to the medical equipment company or to Devon Medical Products directly Power adapter The extriCARE de...

Страница 17: ...en damaged in transit or has been subject to misuse neglect or accident or has been used in violation of Devon Medical Products instructions including without limitation the instructions contained in...

Страница 18: ...io frequency transmitting equipment and other sources of electrical noise in a healthcare environ ment high levels of interference may induce an abnormal stoppage or other disruption of this device Th...

Страница 19: ...for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5s 3 A m 6 kV contact 8 kV air 2 kV for power supply lines 1 kV line s and neutral 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in U...

Страница 20: ...higher the frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Fie...

Страница 21: ...ance according to frequency of transmitter m output power of transmitter W 0 01 0 1 1 10 100 150 kHz to 80 MHz d 1 2 P 0 12 0 38 1 2 3 8 12 80 MHz to 800 MHz d 1 2 P 0 12 0 38 1 2 3 8 12 800 MHz to 2...

Страница 22: ...50 69 Borquist O Ingemansson R et al Individualizing the Use of Negative Pre sure Wound Therapy for Optimal Wound Healing A focused Review of the Literature Ostomy Wound Management 2011 57 4 44 54 De...

Страница 23: ...Guidelines for Managing Pressure Ulcers with Negative Pressure Wound Therapy Advances in Skin and Wound Care 2004 17 1 16 Kirby M Negative Pressure Wound Therapy British Journal of Diabetes Vascular D...

Страница 24: ...www devonmedicalproducts com 866 446 0092...

Отзывы: