After chamber 4 the screen will read EXIT.
Push the SHUTTLE KNOB to confirm.
Access the second parameter, TREATMENT
TIME, by turning the SHUTTLE KNOB to the right.
Push the SHUTTLE KNOB to select TREATMENT
TIME.
You can increase the treatment time by turning
the SHUTTLE KNOB to the right, and decrease
time by turning to the left (treatment time can
range from 15 – 180 minutes, increments of 15
minutes).
Push the SHUTTLE KNOB to confirm the new
treatment time.
Access the third parameter, CYCLE TIME, by
turning SHUTTLE KNOB to the right. Push the
SHUTTLE KNOB to select CYCLE TIME.
You can increase the cycle time by turning the
SHUTTLE KNOB to the right, and decrease time
by turning to the left (cycle time can range from
20 – 80 seconds, increments of 10 seconds).
Push the SHUTTLE KNOB to confirm the new
cycle time.