Access the second parameter, GRADIENT,
by turning the SHUTTLE KNOB to the right.
Push the SHUTTLE KNOB to select GRADIENT.
You can increase the gradient by turning the SHUT-
TLE KNOB to the right, and decrease by turning to
the left (gradient can range from 5 – 15%, incre-
ments of 1%).
Push the SHUTTLE KNOB to confirm the new
gradient.
Access the third parameter, TREATMENT TIME,
by turning the SHUTTLE KNOB to the right.
Push the SHUTTLE KNOB to select TREATMENT
TIME.
You can increase the treatment time by turning the
SHUTTLE KNOB to the right, and decrease time by
turning to the left (treatment time can range from
15 – 180 minutes, increments of 15 minutes).
Push the SHUTTLE KNOB to confirm the new
treatment time.
Access the fourth parameter, CYCLE TIME, by
turning SHUTTLE KNOB to the right. Push the
SHUTTLE KNOB to select CYCLE TIME.