Attach left and/or right quick connectors of the
garment to the quick connect ports which are located
on the front panel of the device. During a single gar-
ment session (i.e. left only or right only) insert the air
blocker into the unused quick connect port.
Putting the garment on:
a) For leg garments, unzip the garment all the
way to the end. Place the foot at the bottom of the
garment and pull up the zipper while supporting the
garment to wrap around the leg.
b) For arm garments, slide the arm through the
internal cavity of the garment.
Press main POWER SWITCH up to the ON
position which is located on the rear panel.
The green power indicator on the front panel
will then illuminate.
Upon this display, User can directly push START
to run the device on FACTORY DEFAULT mode.
• If GRADIENT MODE is desired, continue
to steps 10 - 29.
• If PRESSURE MODE is desired, proceed to
steps 30 - 46.
04
06
07
08
09
06
07
09
THERAPY SETUP
(Settings can only be modified or restored before or in between treatments.)
There are three operating modes with this device:
• Factory Default Mode: Device comes with a factory default setting providing 45mmHg pressure on distal
chamber and a 7% of gradient pressure, and a 30 minute treatment time with a 30 second cycle time, with
no chambers being skipped.
• Gradient Mode: This mode allows you to set your starting pressure
and the percentage drop of pressure gradient between chambers.
• Pressure Mode: This mode allows you to set the pressure in each of
the four chambers in the sleeve.
Follow steps 6 – 8 before pressing the green START button.