F-5
95-8676
3.1
General Status Bits
These bits are used to signal the current operating mode of the device.
Eclipse Status Information (continued)
Description
Address
Value
Min Temperature
40239
Float LSW
40240
Float MSW
Min Temp Hour
40241
Unsigned Long LSW
40242
Unsigned Long MSW
Min Temp (Since Reset)
40243
Float LSW
40244
Float MSW
Min Temp Hour (Since Reset)
40245
Unsigned Long LSW
40246
Unsigned Long MSW
Fixed 4 to 20 mA Value
40247
Float LSW
40248
Float MSW
Reserved
40249
Reserved
40250
Reserved
40251
Reserved
40252
Zero Ratio
40253
Float LSW
40254
Float MSW
Span Factor
40255
Float LSW
40256
Float MSW
5 Volt Power Supply Value
(As read by ADC)
40257
Float LSW
40258
Float MSW
12 Volt Power Supply Value
(As read by ADC)
40259
Float LSW
40260
Float MSW
24 Volt Power Supply Value
(As read by ADC)
40261
Float LSW
40262
Float MSW
Name
Bit
Description
Device Fault (any fault)
0
Set for all fault conditions
Calibration Active
1
Set during calibration
Warm up Mode
2
Set during warm-up
Low Alarm Active
3
Set while alarm is active
High Alarm Active
4
Set while alarm is active
Output Current Fixed
5
Set when output current is fixed
Modbus Write Protect
6
0 = Locked 1 = Unlocked
Calibration Input Active
7
True while the cal line is active
Magnetic Switch Active
8
True while the onboard magnetic switch is active
Hart Initiated Self Test
9
True when self test is initiated from the Hart interface
Reserved
10
Response Test Active
11
True during the gas response test.
Manual Self Test Active
12
True during manual self test