Operation

NOTE: If the Combo Tester X3+ is located near a 

restroom, sink or other water source, operators will need 

to be instructed to thoroughly dry their hands before 

testing. Wet hands may cause inaccurate test results 

and damage to the tester.
1.  A circling light around the test switch indicates when 

the Combo Tester X3+ is on standby and ready to 

perform a test. 

2.  Follow the prompt on the tester’s display. 
3.  When performing a footwear test, be sure to place 

both feet on the dual foot plate (one foot per plate). 

 

NOTE: Keep the foot plate clean with 99% isopropyl 

alcohol when using the 1 Gigohm high test limit. A 

dirty foot plate could yield a false pass. 

 

When performing a wrist strap* test, be sure to 

completely plug in the wrist cord into the tester’s  

jack. 

4.  To begin the test, use your finger to bridge the test 

switch’s inner and outer contacts. The blue standby 

LED will become solid to indicate that the test has 

been initiated. Hold your finger down until the test 

results are displayed on the touchscreen. 

 

If your finger is removed too early, the tester’s blue 

LEDs will blink three times to indicate that the test 

was not completed. DO NOT touch any other metal 

while performing your test as this will affect your 

results. 

5.  The relay terminal will activate if the defined tests 

are passed (if applicable).

NOTE: Failures may be caused by dry skin or minimal 

sweat layer. For wrist straps, try using an approved 

dissipative hand lotion such as 

Menda Reztore™ ESD 

Hand Lotion

 prior to use. Footwear test results can be 

improved by taking a short walk to build a sweat layer 

for better conductivity.
*The Combo Tester X3+ may also be used to 

test smocks or garments that feature a grounding 

mechanism for operators using a coiled cord 

connection.

Maintenance

To maintain optimum performance, cleaning should 

be performed on a regular basis. Use 99% isopropyl 

alcohol to clean the foot plate and test switch. Other 

cleaners are susceptible to leaving residue on these 

surfaces.
NOTE: Avoid wetting or mopping the foot plate. This 

may yield false test results and cause damage to its 

internal circuitry.

Figure 12. Performing a test with the Combo Tester X3+

Содержание Combo Tester X3+

Страница 1: ...t test signal increases test reliability Pairs with the EMIT 50755 ESD Glove Test Fixture to test operators that wear ESD gloves or finger cots per Compliance Verification required by ESD TR53 September 2019 Figure 1 Desco ASK 37525 Combo Tester X3 The factory test limits are set to Wrist Straps 750 kilohms and 10 megohms Footwear 750 kilohms and 35 megohms See page 5 for alternate test limit sett...

Страница 2: ...mployee Owned DESCO WEST 3651 Walnut Avenue Chino CA 91710 909 627 8178 DESCO EAST One Colgate Way Canton MA 02021 1407 781 821 8370 Website Desco com Features and Components Figure 2 Combo Tester X3 features and components A B C D E F G H I J K L M ...

Страница 3: ...terminal termination of the cord to equipment ground This connection will remove any static charge from the user before the test NOTE Failure to correctly ground the Combo Tester X3 may result in damage not covered under warranty M ESD Glove Test Fixture Port Used for connecting the EMIT 50755 ESD Glove Test Fixture See TB 6586 for more information Installation Hardware Setup If the Combo Tester X...

Страница 4: ...bly into the single wire jack on the face of the Combo Tester X3 Be sure to clear the clip at the base of the plunger s shaft when inserting into the jack 6 Connect the Combo Tester X3 to the bracket Use the included thumbscrew to secure the tester as shown below Figure 4 Securing the Combo Tester X3 to the mounting bracket Figure 5 Use the EMIT 50416 Stand as a mounting alternative 2 The plunger ...

Страница 5: ...50 Kilohms 7 5 x 105 HIGH Limit Resistance HIGH Limit Resistance 10 Megohms 1 0 x 107 10 Megohms 1 0 x 107 35 Megohms 3 5 x 107 35 Megohms 3 5 x 107 100 Megohms 1 0 x 108 1 Gigohm 1 0 x 109 Default Settings A dirty foot plate could result in a false pass when the footwear limit is set to 1 Gigohm Be sure to keep the foot plate clean with 99 isopropyl alcohol when using this setting This setting is...

Страница 6: ...ill become solid to indicate that the test has been initiated Hold your finger down until the test results are displayed on the touchscreen If your finger is removed too early the tester s blue LEDs will blink three times to indicate that the test was not completed DO NOT touch any other metal while performing your test as this will affect your results 5 The relay terminal will activate if the def...

Страница 7: ...r Power Input 100 240VAC 50 60 Hz Power Output 5VDC 3 0A Cable Length 5 ft 1 5 m Operating Temperature 70 F to 85 F 21 C to 30 C for 1 gigohm test limit 41 F to 85 F 5 C to 30 C for all other test limits Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C decreasing linearly to 50 85 F 30 C Maximum relative humidity of 50 a...

Отзывы: