TB-6601

 Page 7 of 7

© 2019 DESCO INDUSTRIES INC

Employee Owned

DESCO WEST 

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178

DESCO EAST 

- One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website: 

Desco.com

Calibration

Frequency of recalibration should be based on the 

critical nature of those ESD sensitive items handled 

and the risk of failure for the ESD protective equipment 

and materials. In general, Desco recommends that 

calibration be performed annually.
Use the EMIT 50424

 

Limit Comparator and EMIT 50784 

5-Pound Electrode to perform periodic verification (once 

every 6-12 months) of the Combo Tester X3+. The Limit 

Comparator can be used to check the test limits of the 

Combo Tester X3+ without removing it from the factory 

floor.
See 

TB-6581

 for more information.

NOTE: Calibration is not required for the temperature 

and humidity sensor embedded within the Combo 

Tester X3+.

Specifications

Combo Tester

 X3+

Input Voltage and 

Frequency

(External Adapter)

AC/DC Power Adapter
Power Input: 

100-240VAC, 50/60 Hz
Power Output: 

5VDC, 3.0A
Cable Length: 

5 ft. (1.5 m)

Operating Temperature 70°F to 85°F (21°C to 30°C) 

for 1 gigohm test limit
41°F to 85°F (5°C to 30°C) 

for all other test limits

Environmental 

Requirements

Indoor use only at altitudes 

less than 6500 ft. (2 km)
Maximum relative humidity 

of 80% up to 85°F (30°C) 

decreasing linearly to 50% @ 

85°F (30°C)
Maximum relative humidity of 

50% at 1 gigohm setting

Dimensions

6.75" x 5.00" x 1.75"

(17.1 cm x 12.7 cm x 4.4 cm)

Weight

1.8 lbs (0.8 kg)

Test Accuracy

±20% for 1 gigohm footwear 

test limit
±10% for all other test limits

Test Switch Voltage

5 VDC @ open circuit

Wrist Strap and 

Footwear Test Voltage

30 VDC @ open circuit
Test current is limited by 

resistors and varies on 

the test range setting (100 

kilohms - 1 gigohm)

Temperature Accuracy ±0.5°C 
Humidity Accuracy

±5%

Dual Independent Foot Plate

Dimensions

13.25" x 15.25" x 1.125"

(337 mm x 387 mm x 29 mm)

Weight

5.5 lbs (2.5 kg)

Figure 13. EMIT 50424 Limit Comparator

Figure 14. EMIT 50784 5-Pound Electrode for Limit 

Comparator

Limited Warranty, Warranty Exclusions, Limit 

of Liability and RMA Request Instructions

See the Desco Warranty - 

Desco.com/Limited-Warranty.aspx

Содержание Combo Tester X3+

Страница 1: ...t test signal increases test reliability Pairs with the EMIT 50755 ESD Glove Test Fixture to test operators that wear ESD gloves or finger cots per Compliance Verification required by ESD TR53 September 2019 Figure 1 Desco ASK 37525 Combo Tester X3 The factory test limits are set to Wrist Straps 750 kilohms and 10 megohms Footwear 750 kilohms and 35 megohms See page 5 for alternate test limit sett...

Страница 2: ...mployee Owned DESCO WEST 3651 Walnut Avenue Chino CA 91710 909 627 8178 DESCO EAST One Colgate Way Canton MA 02021 1407 781 821 8370 Website Desco com Features and Components Figure 2 Combo Tester X3 features and components A B C D E F G H I J K L M ...

Страница 3: ...terminal termination of the cord to equipment ground This connection will remove any static charge from the user before the test NOTE Failure to correctly ground the Combo Tester X3 may result in damage not covered under warranty M ESD Glove Test Fixture Port Used for connecting the EMIT 50755 ESD Glove Test Fixture See TB 6586 for more information Installation Hardware Setup If the Combo Tester X...

Страница 4: ...bly into the single wire jack on the face of the Combo Tester X3 Be sure to clear the clip at the base of the plunger s shaft when inserting into the jack 6 Connect the Combo Tester X3 to the bracket Use the included thumbscrew to secure the tester as shown below Figure 4 Securing the Combo Tester X3 to the mounting bracket Figure 5 Use the EMIT 50416 Stand as a mounting alternative 2 The plunger ...

Страница 5: ...50 Kilohms 7 5 x 105 HIGH Limit Resistance HIGH Limit Resistance 10 Megohms 1 0 x 107 10 Megohms 1 0 x 107 35 Megohms 3 5 x 107 35 Megohms 3 5 x 107 100 Megohms 1 0 x 108 1 Gigohm 1 0 x 109 Default Settings A dirty foot plate could result in a false pass when the footwear limit is set to 1 Gigohm Be sure to keep the foot plate clean with 99 isopropyl alcohol when using this setting This setting is...

Страница 6: ...ill become solid to indicate that the test has been initiated Hold your finger down until the test results are displayed on the touchscreen If your finger is removed too early the tester s blue LEDs will blink three times to indicate that the test was not completed DO NOT touch any other metal while performing your test as this will affect your results 5 The relay terminal will activate if the def...

Страница 7: ...r Power Input 100 240VAC 50 60 Hz Power Output 5VDC 3 0A Cable Length 5 ft 1 5 m Operating Temperature 70 F to 85 F 21 C to 30 C for 1 gigohm test limit 41 F to 85 F 5 C to 30 C for all other test limits Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C decreasing linearly to 50 85 F 30 C Maximum relative humidity of 50 a...

Отзывы: