Relay Terminal

The Combo Tester X3+ features a relay terminal that 

can be integrated with electronic door locks, lights, 

buzzers, etc. to control access to an area.
The relay open and close (activation) time may be 

modified in the Admin menu of the tester’s software.

Contact Rating

1A @ 30VDC, .5A @ 

125VAC

Maximum Switching Power 30W
Maximum Switching Voltage 250VAC, 220VDC
Maximum Switching Current 1A

Test Limit Configuration

The following footwear and wrist strap resistance limits 

are available in the Combo Tester X3+:

Footwear

Wrist Strap

LOW Limit Resistance

LOW Limit Resistance

100 Kilohms (1.0 x 10

5

)

100 Kilohms (1.0 x 10

5

)

750 Kilohms (7.5 x 10

5

)

750 Kilohms (7.5 x 10

5

)

HIGH Limit Resistance

HIGH Limit Resistance

10 Megohms (1.0 x 10

7

)

10 Megohms (1.0 x 10

7

)

35 Megohms (3.5 x 10

7

)

35 Megohms (3.5 x 10

7

)

100 Megohms (1.0 x 10

8

)

1 Gigohm (1.0 x 10

9

)**

Default Settings

**A dirty foot plate could result in a false pass when the 

footwear limit is set to 1 Gigohm. Be sure to keep the 

foot plate clean with 99% isopropyl alcohol when using 

this setting. This setting is not recommended for relative 

humidity greater than 50%.

Normally Closed
Common
Normally Open

Figure 8. Relay terminal contacts located on the back of 

the Combo Tester X3+

SETTING THE RESISTANCE LIMITS

1.  Access the Administrator Menu by entering an 

administrator ID number and PIN on the keypad. 

2.  Tap the Preferences button. 

Figure 9. Locating the Preferences button in the Admin 

Menu

Figure 10. Locating the ESD Test button in the 

Preferences Menu

3.  Tap the ESD Test button. 

4.  Adjust the Wrist Strap and Footwear Limits by 

tapping the < and > buttons. Tap the Save button 

when complete.

Figure 11. Locating the Wrist and Foot Limits in the ESD 

Test Menu

Содержание Combo Tester X3+

Страница 1: ...t test signal increases test reliability Pairs with the EMIT 50755 ESD Glove Test Fixture to test operators that wear ESD gloves or finger cots per Compliance Verification required by ESD TR53 September 2019 Figure 1 Desco ASK 37525 Combo Tester X3 The factory test limits are set to Wrist Straps 750 kilohms and 10 megohms Footwear 750 kilohms and 35 megohms See page 5 for alternate test limit sett...

Страница 2: ...mployee Owned DESCO WEST 3651 Walnut Avenue Chino CA 91710 909 627 8178 DESCO EAST One Colgate Way Canton MA 02021 1407 781 821 8370 Website Desco com Features and Components Figure 2 Combo Tester X3 features and components A B C D E F G H I J K L M ...

Страница 3: ...terminal termination of the cord to equipment ground This connection will remove any static charge from the user before the test NOTE Failure to correctly ground the Combo Tester X3 may result in damage not covered under warranty M ESD Glove Test Fixture Port Used for connecting the EMIT 50755 ESD Glove Test Fixture See TB 6586 for more information Installation Hardware Setup If the Combo Tester X...

Страница 4: ...bly into the single wire jack on the face of the Combo Tester X3 Be sure to clear the clip at the base of the plunger s shaft when inserting into the jack 6 Connect the Combo Tester X3 to the bracket Use the included thumbscrew to secure the tester as shown below Figure 4 Securing the Combo Tester X3 to the mounting bracket Figure 5 Use the EMIT 50416 Stand as a mounting alternative 2 The plunger ...

Страница 5: ...50 Kilohms 7 5 x 105 HIGH Limit Resistance HIGH Limit Resistance 10 Megohms 1 0 x 107 10 Megohms 1 0 x 107 35 Megohms 3 5 x 107 35 Megohms 3 5 x 107 100 Megohms 1 0 x 108 1 Gigohm 1 0 x 109 Default Settings A dirty foot plate could result in a false pass when the footwear limit is set to 1 Gigohm Be sure to keep the foot plate clean with 99 isopropyl alcohol when using this setting This setting is...

Страница 6: ...ill become solid to indicate that the test has been initiated Hold your finger down until the test results are displayed on the touchscreen If your finger is removed too early the tester s blue LEDs will blink three times to indicate that the test was not completed DO NOT touch any other metal while performing your test as this will affect your results 5 The relay terminal will activate if the def...

Страница 7: ...r Power Input 100 240VAC 50 60 Hz Power Output 5VDC 3 0A Cable Length 5 ft 1 5 m Operating Temperature 70 F to 85 F 21 C to 30 C for 1 gigohm test limit 41 F to 85 F 5 C to 30 C for all other test limits Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C decreasing linearly to 50 85 F 30 C Maximum relative humidity of 50 a...

Отзывы: