background image

TB-3043 Page 4 of 5

© 2009 DESCO INDUSTRIES, INC.

Employee Owned

Compliance Verification

ANSI/ESD S20.20-2007 added Compliance Verification columns 

to the Tables. The test method for Compliance Verification 

is ESD TR53. Ionizers should be tested periodically for both 

polarity’s discharge times, and for offset voltage balance per ESD 

Association Technical Report ESD TR53 Compliance Verification 

of ESD Protective Equipment and Materials. The most accurate 

tool to use for this is a Charged Plate Monitor. Alternatively, a 

portable battery operated Ionization Test Kit can be used.

Per ESD TR 53 ANNEX A “Test Frequency, The objective of the 

periodic test procedures listed in this document is to identify if 

significant changes in ESD equipment and materials performance 

have occurred over time.  Test frequency limits are not listed in 

this document, as each user will need to develop their own set of 

test frequencies based on the critical nature of those ESD sensitive 

items handled and the risk of failure for the ESD protective 

equipment and materials.”

Per ESD TR53, all the test locations of S3.1 are not required; 

rather “Measurements should be made at the location where ESD 

sensitive items are to be ionized. For many EPA ESD Control 

items sampling is appropriate for Compliance Verification, however, 

best practice is to test each ionizer.

NIST Calibration

Desco provides a basic, National Institute of Standards and 

Technology (NIST) traceable calibration for the products that 

we manufacture.  This is sometimes referred to as a Level 1 

calibration.  

For more on National Institute of Standards and Technology see:

http://www.nist.gov/index.html

For more information on the calibration that Desco’s provides 

for products that we manufacture see:  

http://www.desco.com/

Calibration.aspx

 

Calibration / Offset Voltage Balance Adjustment

When an alarm sounds, most users will clean emitter pins (see 

Maintenance / Alarms section) and calibrate the ionizer. Per ESD 

TR 53 section 5.3.6.7.1 “The best practice is to measure the offset 

voltage and discharge times, clean the unit, including emitter points 

and air filters if present, offset voltage to zero (if adjustable), and 

then repeat offset voltage and discharge time testing. If the unit 

does not meet offset voltage specifications or minimum established 

discharge time limits, further service is indicated. Manufacturers 

should provide details on service procedures and typical service 

intervals.”

JESDD625-A (Revision of EIA-625) recommends ionizers be 

tested semiannually, noting to use “S3.1 except the number of 

measurement points and locations may be selected based on the 

application.”

NOTE

: A charged plate analyzer or monitor should be used in 

order to properly calibrate the Chargebuster High Outfit Benchtop 

Ionizer.EMIT offers the 

50555

 Charged Plate Analyzer.

I. Properly setup the ionizer as described in the Installation  

   procedure on page 1.

 

II. Turn the unit ON and set the FAN SPEED to HIGH.

III. Position the charged plate analyzer 18 inches away from High 

     Output Benchtop Ionizer.

IV. The offset voltage balance should be within 0 ±15 volts. The  

       

     required limit per ANSI/ESD S20.20 is less that ± 50 volts.     

     To increase the output in a postive direction, turn the BALANCE 

     ADJUST potentiometer in a clockwise direction.  To increase 

     the output in a negative direction, turn the BALANCE ADJUST 

     potentiometer in a counter clockwise direction. 

V. Test the neutralization discharge time by applying a ±1,000 

     volt on the charged plate. Do both ± polarities.  The      

     neutralization (discharge) time should be less than 3 seconds.  

     See figures 4 for typical discharge times. The required limit per  

     ANSI/ESDS20.20 is “user defined”.

VI. Submit the offset voltage balance to the ionizer’s control 

      circuit by quickly pressing the ALARM RESET button. The 

      STATUS LED should turn off and then illuminate green to   

      verify that the control circuit was successfully programmed.

 

VII.Test each fan’s alarm by shorting its two grills located on the 

      bottom side of the ionizer (see Figure 8). The alarm should 

      sound and the STATUS LED should illuminate red.

Health

There are no known health risks associated with our devices. The 

emitters work at about 4-6 kV and can create ozone, but there 

have been no significant measurement of ozone from our emitter 

sets as all our existing units test well below the OSHA limit of 0.05 

ppm ozone.  For additional safety information, see “Dispelling an 

Old Myth” written by William Metz of Hewlett-Packard published in 

Evaluation Engineering magazine September 2001. 

Figure 7. 

60505

 Shorting the Ionizer’s two fan grills

© 2010 DESCO INDUSTRIES INC.

Employee Owned

Содержание 60505

Страница 1: ...c dissipative materials and personnel A complete static control program must also deal with isolated conductors that cannot be grounded insulating materials e g most common plastics and moving personnel who cannot use wrist or heel straps or ESD control flooring and footwear Air ionization is not a replacement for grounding methods It is one component of a complete static control program Ionizers ...

Страница 2: ...t will conduct a self test The audible alarm will sound and then the LED will cycle through RED YELLOW and GREEN The LED will remain GREEN during normal operation Figure 4 Front and Back High Output Benchtop Ionizer Angle Adjustment Knobs The angle of the fan may be adjusted to direct the airflow in the optimum coverage area Visual Alarm GREEN Normal Operation RED Ionizer needs maintence Voltage s...

Страница 3: ...ntification method for each ionizer in your Maintence Calibration schedule You may choose to use the ionizer s serial number for identification as it is a unique number for identification as it is a unique number that can be traced to when the ionizer was manufactured Neutralization Discharge Times The comparative efficiency of overhead ionizers is determined by a standard test published by ANSI E...

Страница 4: ...ge time testing If the unit does not meet offset voltage specifications or minimum established discharge time limits further service is indicated Manufacturers should provide details on service procedures and typical service intervals JESDD625 A Revision of EIA 625 recommends ionizers be tested semiannually noting to use S3 1 except the number of measurement points and locations may be selected ba...

Страница 5: ...ng instructions and ship your unit freight prepaid Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warranty will not apply to defects or damage due to accidents neglect misuse alterations operator error or failure to p...

Отзывы: