SAFETY INSTRUCTIONS
1.1
Intended use
The light Dmed® OPTICLUX is an examination light.
Its intended use is to illuminate the body of the patient to
support diagnosis or treatment. An interruption of the
diagnosis or treatment caused by a light failure is always
possible without posing any risk for the patient. The light
cannot be used in operating rooms.
Moreover, Dmed® OPTICLUX can also be used in the
laboratory or in quality control
.
1.2
User profiles
Medical professional
A medical professional is any person who has completed
medical training and is working in his or her professional
field.
Cleaning professional
A cleaning professional is trained in national and job-
related hygiene regulations.
Electrician
An electrician is trained in the fields of electronics and
electrical engineering and is familiar with the relevant
standards and regulations.
Qualified professional
A qualified professional is capable of mounting and
dismounting the light thanks to professional training,
knowledge and experience and knowledge of the
regulations.
1.3
Safety instructions
Operation by a medical professional
The instructions form part of the product. They must
be stored and made accessible to all subsequent
users.
Any work on the light (including repairs) must be
carried out by qualified electricians only. Mounting
must be performed by a qualified professional only.
The light must not be altered or manipulated in any
way. Only approved original parts must be used.
Any use other than the intended use with original
parts may give rise to other technical values and
life-threatening danger.
Operation
in
potentially
explosive
areas
is
prohibited. The light power supply is a potential
ignition source.
The light must only be operated in dry, dust-free
rooms.
The light must not be left switched on without
supervision.
For lights in protection class I, it is essential for the
earthing conductor to be connected to the light
housing.
Do
not
use
any
light
that
is
damaged.
Defective cables are also potential hazards. Do not
place the cable close to any heat source or on sharp
edges.
Damage to the eyes. Never look directly into the
light source.
Replace any glass that is damaged before operating
the light again.
In order to prevent electric shocks, only connect the
light to a power supply with an earth conductor (only
C-version).
Luminaires which are equipped with a hospital
grade plug (UL 817, only US- and Canada version),
the earth connector is used as a functional earth
within the luminaire.
Never place additional loads on the light head or the
arm.
The light must not be covered by a cloth or any
similar item while it is in operation.
The ventilation openings (where they exist) must be
free whenever the light is in operation.
The light must not be operated near to external heat
sources that exceed the maximum ambient
temperature of the light.
The light must not be used in environmental
conditions other than those for which it is intended.
The light must only be used for the intended use
described in this document.
The manufacturer cannot be held responsible for
any injury or damage that is caused as a result of
any use other than the intended use or of any failure
to comply with safety instructions and warnings.
Do not look into the UV rays (Wood light version
only).
A
fter use, always cover the magnifier with the
magnifier cover (danger of fire or burns).
During an examination in which the eyes may come
into contact with UV light (Wood light version only),
the user must advise the patient to keep his or her
eyes closed.
The lamp is intended for a lifetime of 10 years
1.4
Warning levels
DANGER
Indication of hazards that can lead
to death or serious
injury
if measures are disregarded.
WARNING
Indication of hazards that can lead to
injury
if
measures are disregarded.
CAUTION
Indication of hazards that can lead to
damage to
property
if measures are disregarded.