30
4
Maintenance
4.1
Cleaning and Care
4.1.1
General Information
CAUTION!
Health risks due to the handling of infectious or pathogenic
germs.
Infectious & pathogenic germs in the wound exudate cause health risks.
•
Wear appropriate disposable gloves when replacing the
disposable exudate canister or any tubing.
•
Use of the disposable exudate canister or any tubing is for one
patient only.
•
Replace the disposable exudate canister(s) in accordance to the
respectively applicable hygiene instructions, if it is full, prior to
each new patient or every 3-5 days.
•
For each new patient, ensure unit has been cleaned following
established facility protocols.
•
Components that have come into contact with the wound exudate
must be cleaned, disinfected or disposed of after each use.
•
The disposal of wound exudate and contaminated components
must be performed in a proper manner.
CAUTION!
Health risks due to the handling of disinfectants.
•
The use of appropriate disposable gloves during disinfection is
recommended.
•
Follow the manufacturer’s disinfectant instructions.
CAUTION!
Possible bodily injury by electric shock.
Prior to cleaning and/or disinfection, switch off the device, disconnect the
power supply by unplugging it from the wall and then disconnect the
power supply from the unit of the PRO-II
®
or PRO-III
®
device.
ATTENTION!
Risk of damage to the device due to improper cleaning agents.
•
Do not use disinfectants that contain acetone. These may damage
or disfigure the housing components and the accessories.
•
Follow the instruction for use provided by the manufacturers of
the utilized disinfectants, particularly with respect to material and
surface compatibility as well as the concentration information.
•
DeRoyal recommends Incidin LIQUID for wipe and spray
disinfection of the device.
Disinfection is not mandatory if the device is used for one patient only (in in-home
care). Disinfection is mandatory if used in an in-patient setting and between each
new patient use: in-home or in-patient.
4.1.2
Cleaning and disinfection of the surface of the device
•
Clean and disinfect the surfaces of the device at least once a week or in
accordance to the respectively applicable facility protocols.
•
The devices may be wiped with a damp, lint-free cloth.
•
Follow the information on wipe disinfection provided in the chapter 4.1.1.
•
Never clean or disinfect the device when turned on.
Minor discolorations may occur on the plastic housing components after
prolonged use. These do not affect the function of the device.
If the interior of the device comes into contact with liquid, the device must be inspected by
DeRoyal or an authorized service partner of DeRoyal.
31
4.1.3
Disposal of the disposable exudate canister incl. the suction tube for PRO-
II
®
Device
1.
Close the suction tube of the disposable exudate canister by using the closure
cap of the suction tube.
2.
Dispose of the disposable exudate canister incl. suction tube in a proper
manner. (Please refer to chapter 6.3). It is a single-use item.
4.1.4
Disposal of the disposable canister and the suction tube for PRO-III
®
Device
1.
Close the tubing clamp at the suction tube (fig. 3).
2.
Separate the connecting tube of the canister from the device.
3.
Remove the disposable canister and tubing from device.
4.
Dispose of the disposable canister and tubing in a proper manner. (Please
refer to chapter 6.3 “Disposal”).
4.2
Maintenance and Service
The PRO-II
®
and PRO-III
®
devices are maintenance free if used according to the instruction
for use.
Perform a visual and functional inspection prior to each use. (Please refer to chapter 5.1).
Also include the accessories of the device in the inspection.
Opening of the device and repairs must only be performed by DeRoyal or by authorized
professionals of DeRoyal in compliance with the service documentation specified by the
manufacturer as well as with technical and hygienic precautionary measures.
The device may be sent back for repair to DeRoyal directly or via the specialty dealer from
which the device was purchased.
Clean and disinfect all accessories prior to returning the device. The device itself must be
treated with a surface disinfectant. Please note the guidelines referring to decontamination
before shipping (chapter 6.1).
DeRoyal neither ensures proper functioning of the PRO-II
®
and PRO-III
®
medical suction
devices nor is DeRoyal liable for property damage or personal injury if
-
original DeRoyal accessories or spare parts are not used,
-
the user instructions in this instruction for use have not been followed,
-
installation, settings, modifications, enhancements and repairs are not performed by
DeRoyal or authorized personnel of DeRoyal,
-
the safety seal has been removed or is damaged,
-
or any other portion of the warranty program or user manual are violated, omitted
or disregarded in any way.
4.3
Testing the PRO-II
®
or PRO-III
®
Devices
DeRoyal offers its partners and customers fast and proper testing services.