16
2.2.5
Information on the double filter system for PRO-III
®
Device
The double filter system consists of the external bacterial filter integrated in the canister
tubing and the internal filter installed in the device.
The external bacterial filter is incorporated in the available connection tubing provided by
DeRoyal.
The double filter system effectively protects against overflow and contamination of
the interior of the device.
Service life and reuse
To ensure consistent performance, the internal filter must be replaced
after
contact with the exudate (blocked), after the filter service life has expired
(8,000 hours)
(
symbol in the display) or during
maintenance/repair
.
The internal filter must be replaced by DeRoyal or an authorized service partner of
DeRoyal.
2.2.6
Battery
The charge level of the battery is shown in the display.
It is strongly recommended to fully charge the battery prior to first startup of the PRO-II
®
and PRO-III
®
devices and to repeat this after the each use.
The PRO-II
®
and PRO-III
®
devices are equipped with a lithium-ion battery, which, unlike
traditional types of rechargeable batteries, have a low self-discharge rate.
The PRO-II
®
and PRO-III
®
devices should ideally be stored and charged at room
temperature in accordance with the ambient conditions specified in the technical data.
Never store the device incl. battery in a discharged state.
Fully recharge the battery if the device is not operated for a longer period of time (approx.
10 months).
Lithium-ion rechargeable batteries do not have a memory effect. They can, therefore, be
recharged at any time after initial charging. Only frequent short-time charging should be
avoided.
The batteries of the PRO-II
®
and PRO-III
®
devices are protected against deep discharge,
but the charging information listed above must nevertheless be followed. The batteries also
are protected against overheating during charging. If the battery temperature is exceeded
during charging due to improper ambient conditions, charging is temporarily discontinued
to allow for cooling. The purpose of this measure is to ensure safe operation and to protect
the battery.
The operational service life of the battery is 2 years. According to the manufacturer of the
battery, the battery has a remaining capacity of more than 80% after 300 charge cycles.
17
2.2.7
Pressure settings
Once the PRO-II
®
or PRO-III
®
device has been switched on, the pressure settings can be
individually adjusted by a healthcare professional.
The pressure settings can be adjusted in a range from -20 mmHg to -200 mmHg (in steps
of 5 mmHg). The setting of the pressure and time values is described in chapter 3.2.
Adjustments to device settings must only be made if instructed to do so and only
by healthcare professionals. Prior to switching on the PRO-II
®
or PRO-III
®
device it
must be ensured that the device is equipped with a disposable exudate canister.
2.3
Warranty
These devices are warranted to be free of defects in material and workmanship for one
year from date of shipment. All operational aspects (mechanical and electrical) are
covered by this warranty. The warranty does not cover accessories, spare parts and
consumables. Batteries are covered by a warranty of 6 months from purchase date.
DeRoyal warranty is neither extended nor renewed by warranty/service work.
TO THE EXTENT ALLOWED BY LAW, DEROYAL’S WRITTEN WARRANTIES ARE GIVEN IN
LIEU OF ANY IMPLIED WARRANTIES, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE.
CAUTION: Warranty claims will not be accepted if (1) the unit has been opened or repaired
by unauthorized individuals, (2) the unit has not been used in accordance with proper
guidelines; this includes using any non-DeRoyal provided accessories (kits, canisters, etc.)
(3) the special tamper proof seal is missing or broken, or (4) the unit has been rendered
inoperable due to physical damage (inclusive of drops or falls of the device). Any
maintenance or service repair that is not covered by the warranty may be performed by
DeRoyal at a cost of $75 per hour. A minimum of $75 may be charged and billed
appropriately. Any repair charges less than $500.00 will be automatically billed, unless
otherwise instructed by customer at time service is requested.
DeRoyal is responsible for impacts on safety, reliability and specified performance only if:
•
original DeRoyal accessories and spare parts are used,
•
maintenance and repair are performed by professionals authorized by DeRoyal. or
by DeRoyal itself,
•
the affected product is used and operated in accordance with the instruction for use
and within its intended use.
DeRoyal does not warrant accurate function of the devices PRO-II
®
and PRO-III
®
devices
and is not liable for a loss of property or personal injury in the following circumstances:
•
non-original DeRoyal accessories or spare parts are used,
•
use information of this instruction for use are ignored,
•
installation, settings, changes, upgrades or repairs are not carried out by DeRoyal or
by professionals authorized by DeRoyal,
•
the safety seal is broken or removed.
All warranty claims are voided if the device is opened by unauthorized individuals, the
safety seal is removed / damaged or repairs have been performed by unauthorized
individuals.
Содержание Prospera PRO Series