26
16.
You get back to the overview screen.
17.
Switch off the PRO-II
®
or PRO-III
®
devices by pressing the
button for 3 seconds.
3.2.4
Language selection
The PRO-II
®
and PRO-III
®
devices allow the user to select
a
language
at initial startup. The
selected language is stored and automatically loaded at each startup. To customize the
language, follow these steps:
1.
Press the
button for 1-2 seconds to switch on the PRO-II
®
or PRO-III
®
devices. The
following start screen is displayed for 5 seconds:
2.
While the start screen is displayed, simultaneously press the
arrow buttons. The
menu
Setup
is displayed.
3.
Use the
arrow buttons to select the
Language
menu.
4.
Use the
button to confirm your choice.
5.
Use the
arrow buttons to select the desired language:
6.
Use the
button to confirm your choice.
3.2.5
Alarm delay at startup
At startup of the PRO-II
®
or PRO-III
®
device, the alarm “System closed” for
kinked tubing or too tight wound dressing and the alarm “Check dressing seal”
are delayed for 60 seconds to ensure that the wound dressing can be placed
without activating alarms.
DeRoyal
27
3.3
Patient Mode
The PRO-II
®
and PRO-III
®
devices
allow the user to select
patient mode
at startup. In
patient mode, the settings for the alarm sensitivities “System closed” and “Check dressing
seal” as well as the patient runtime can be viewed and reset. To select the patient mode,
follow these steps:
1.
Press the
button for 1-2 seconds to switch on the PRO-II
®
or PRO-III
®
device. The
following start screen is displayed for 5 seconds:
2.
While the start screen is displayed, press and hold down the
button and additionally
press the
button for 1-2 seconds. The
Authorization
screen is displayed.
3.
Use the
arrow buttons to enter the “1111” code.
Press the
arrow button until the desired digit of the code is displayed and confirm
the entry with the
button. Select the other digits of the code with the
arrow
button and confirm them with the
button as well.
The authorization code may only be passed to specially trained personnel.
You will get the training and the authorization code by DeRoyal or an
authorized distribution partner of DeRoyal.
Passwords must be treated as confidential information to prevent misuse.
4.
After the authorization, the sensitivity setting of the alarm “System closed” is
displayed.
The sensitivity can be adjusted with the
arrow buttons.
1
5
9
low sensitivity
default
high sensitivity
(delayed triggering of the
alarm)
(rapid triggering of the alarm)
When the sensitivity is set to “1,” the device detects a full canister or a
blocked filter only.
For this setting, it is highly recommended to reduce the interval time for
checking the dressing seal by the healthcare personnel to detect potential
blockings in the aspirating system promptly.
5.
Press the
button to confirm the entry.