27
Error code
Error
Cause
iVÉ,ii`Þ
At the
time of
charging
E-C
Low voltage of
batteries
The voltage of
batteries is too
low (The batteries
>ÀiÊy>ÌÊÀÊÌÊ
inserted).
Put the batteries
into the battery
chamber, or replace
with new batteries.
High voltage of
batteries
The voltage of
batteries is too
high (Malfunction
of circuit).
Contact your Technical
Service Department.
E-E
Beyond the range of
working temperature
Beyond the range
of working
temperature
or break in the
thermistor in the
battery section.
Use within the range
of working temperature,
or replace the batteries.
At the
time of
calibra-
tion
CEO
Beyond the upper limit
The operating
life of the motor
handpiece or
contra-angle has
expired.
,i«>ViÊÌ iÊÌÀ
handpiece or contra-angle
head.
CE 1
Below the lower limit
Other
E-F
Abnormal heat
generation from
batteries
The batteries
generate
abnormal heat.
,i«>ViÊÌ iÊL>ÌÌiÀið
If the heat generates from
the new set of batteries,
malfunction of the circuit
may be suspected. Contact
Dentsply Tulsa Dental
Specialties.