ENGLISH • 1
Cavitron
®
Select™ SPS™
Ultrasonic Scaler
Cavitron
®
Select™ SPS™
Ultrasonic Scaler with
Cavitron
®
Select™
Reservoir System
Cavitron
®
Select™ SPS™ Ultrasonic Scaler with Cavitron
®
Select™ Reservoir System
Introduction
Directions For Use
Please read carefully and completely before operating unit.
DENTSPLY
®
Professional is an ISO13485 certiied company. he
Cavitron
®
Select™ System is classiied by Underwriters Laboratories
Inc. with respect to electric shock, ire, mechanical hazards only in
accordance UL 2601-1 1 and Can/CSA C22.2 NO. 601.1.
he Scaling System operates by converting ordinary AC house
current into high frequency current. he ultrasonic system consists
of two parts: an insert, and the SPS™ (Sustained Performance
System) electronic system. he system incorporates two closed
loops. One loop provides automatic tuning (operating frequency
is adjusted to be at resonance for each insert), the second loop
automatically controls the tip stroke over diferent working
conditions. he scaler produces 30,000 microscopically small
strokes per second at the insert’s working tip. his combined with
acoustic efects of the coolant water, produces a synergistic action
that literally “powers away” the heaviest calculus deposits while
providing exceptional operator and patient comfort.
he Pump/Reservoir accessory for the Cavitron
®
Select™ SPS™ is an
optional lavage dispenser. Footswitch plug, handpiece plug, water
tube and power supply plug are removed from the Scaler before
mating with the Pump/Reservoir. he Scaler is snapped together
with the Pump/Reservoir. he footswitch plug, handpiece plug
and power supply plug are inserted into mating receptacles at the
rear of the Pump/Reservoir. Ultrasonic signal, footswitch and DC
power connections between the units are made via an edgecard
connector at the rear of the Scaler and the front of the Pump
module. he Reservoir has a capacity of 450 ml. Fluid from the
Reservoir is propelled from the Reservoir Bottle with 15–19 psig
compressed air generated by an internal diaphragm pump.
ON/OFF of the lavage is controlled through an internal solenoid
valve. Fluid to the Handpiece passes directly from the Pump/
Reservoir to the Handpiece. No luid passes through the Cavitron
®
Select™ SPS™ when used with this accessory. Flow through any
insert in the Handpiece will be greater than 40 cc/m water when
adjusted to maximum.
Technical Support
For technical support and repair assistance in the U.S., call
1-800-989-8826 Monday through Friday, 8:00 AM to 5:00 PM
(Eastern Time). For other areas, contact your local DENTSPLY
®
representative.
Supplies & Replacement Parts
To order supplies or replacement parts in the U.S., contact your
local DENTSPLY
®
Distributor or call 1-800-989-8826
Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time).
For other areas, contact your local DENTSPLY
®
representative.
Caution: U.S. Federal Law restricts this device to sale by or on the
order of a liscense dental professional.
For dental use only.
ENGLISH
Содержание Cavitron Select SPS
Страница 6: ......