6 Follow-up
Operating Instructions Midwest
®
E Pro / Midwest
®
E Plus / Midwest
®
E mini
46
66 04 206 D3686
.
201.01.04.02
6
Follow-up
6.1
After each treatment session
NOTICE! Condition immediately, or at the latest, one hour after
treatment.
NOTICE! Use only Midwest Plus Aerosol Spray.
✔ The bur instrument must not be moving.
✔ Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur instrument with tweezers.
3. Pre-disinfect directly at the treatment center [ → 48].
4. Detach the instrument from the motor.
5. Transport the instrument to the hygiene room in a suitable transport
container.