Operating Instructions Midwest
®
E Pro / Midwest
®
E Plus / Midwest
®
E mini
7 Conditioning
66 04 206 D3686
.
201.01.04.02
49
7.2
Automated cleaning
The cleaning equipment used must be approved by its manufacturer for
the cleaning of dental instruments.
For further details, refer to the operating instructions supplied with the unit.
✔ The instrument is conditioned with a cleaning device.
1. Check whether the instrument is clean after reprocessing under
good lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If they are still dirty, repeat the process.
For further reprocessing, the instrument should be dry and free of residue.
3. Blow the instrument out with max. 3 bar.
4. Maintain mechanical parts manually [ → 52].
5. Maintain the push button chuck manually [ → 54].
6. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Conduct sterilization [ → 58].