3 CEREC Bluecam
3.1 CEREC Bluecam without sapphire glass
Dentsply Sirona
Instructions for CEREC Bluecam and CEREC Omnicam
20
3.1.8
Hot air sterilization of the prismatic tube
NOTE
Hot air sterilization and high-level disinfection must not be combined.
The process for hot air sterilization is as follows:
1.
Remove the camera support (if present).
2.
Carry out a preliminary cleaning process (see "Pre-cleaning of the
prismatic tube [→ 17]").
3.
Remove the prismatic tube from the CEREC Bluecam (see the
section on “Removing the prismatic tube [→ 16]”).
4.
The prismatic tube can be sterilized using hot air (180 °C, 30 min).
Please ask your dealer for deals on hot air sterilizers. Place the
prismatic tube in the hot air sterilizer and follow the manufacturer's
instructions.
5.
Store the prismatic tube in such a way that it is protected from
contamination until the next use.
6.
Carefully re-attach the prismatic tube and allow it to lock in place
(see section on “Refitting the prismatic tube [→ 17]”).